FISHER HAMILTON SAFEAIRE series Скачать руководство пользователя страница 5

SAFEAIRE

®

LABORATORY FUME HOODS

OPERATION, MAINTENANCE AND INSTALLATION INSTRUCTIONS

INSTALLATION

5

Dimensions are nominal and illustrations and specifications are based
on the latest product information available at the time of publication.
The right is reserved to make changes at any time without notice.

1. Attach left-hand side assembly (2) to back assembly (1).

Take back panel assembly and stand vertically, as shown,
and slide into end assembly. (Angle flange of end assembly
fits between back panel and horizontal members of back
assembly). Secure each corner with two #8 X 5/8" PPHSMS
at mating holes on top and bottom corners of joined 
assembly. Repeat steps with right-hand end (3). Attach each
baffle support with two each #8 X 3/4" PPHTCS through
back panel into support. Do not over-tighten.

#8 X3/4" PPHTCS

#8 X5/8" PPHSMS

1

2

1

2

ASSEMBLY INSTRUCTIONS FOR KNOCKED-DOWN
FLOOR-MOUNTED FUME HOODS

General - When receiving fume hood assemblies, inspect for
damage immediately. If damage is noted, request the delivering
carrier to note and describe damage on bill of lading prior to
your  signature. Notify carrier immediately and request 
inspection.

Open boxes and crates and examine for hidden damage. If 
damage is noted, save containers and request an inspection.

Move all components to installation area.

2. Attach top assembly (4), as shown at left, to

back and ends with eleven (11) #10 X 5/8"
PPHSMS. Top assembly fits underneath right-
hand rear pulley bracket assembly. Secure
exhaust collar to rear rail.

#10 X5/8" PPHSMS

2

TOP

BACK

REAR PULLEY
BRACKET ASSEMBLY

Содержание SAFEAIRE series

Страница 1: ......

Страница 2: ...upply Chamber Air Filter 26 Exhaust Filter Replacement 26 Blower RPM Adjustments 27 Manometer Monitoring Exhaust Filters 27 FIELD TESTING 28 30 TROUBLESHOOTING 31 GENERAL Fume hoods are exposed to ext...

Страница 3: ...impact friction or spontaneous combustion Persons using perchloric acid should be thoroughly familiar with its hazards Many reported laboratory accidents have involved less than one gram of reactant...

Страница 4: ...ash hold down clips Open the sash and remove blocking being careful not to damage sill or baffles NOTE Baffles can be installed at this time or after the hood is in place on the work surface See Pages...

Страница 5: ...nd end 3 Attach each baffle support with two each 8 X 3 4 PPHTCS through back panel into support Do not over tighten 8 X3 4 PPHTCS 8 X5 8 PPHSMS 1 2 1 2 ASSEMBLY INSTRUCTIONS FOR KNOCKED DOWN FLOOR MO...

Страница 6: ...embly 6 using three 3 8 X 5 8 PPHSMS Slide 3 sheave 7 then 2 sheave 8 over threaded stud Loosely fasten with 1 4 20 KEPS nut Install right hand cable retainer support 27 using two 2 8 X 5 8 PPHSMS 27...

Страница 7: ...e the same size Insert upper baffle into collar extrusion per Detail B Test remote control baffle adjuster for function 16 15 14 13 12 COLLAR TOP PANEL UPPER BAFFLE DETAIL B INTER MEDIATE BAFFLE LOWER...

Страница 8: ...Make any necessary adjustment If necessary fine tune weight pan Fasten clamp screws tightly Attach all cable retainers 22 25 Do not over tighten retainers allow 1 16 for pulley movement See Detail D...

Страница 9: ...the time of publication The right is reserved to make changes at any time without notice 9 Attach lintel panel 28 to front of hood Engage top of panel with studs on top inside of front access panel a...

Страница 10: ...he side returns of the Top Filler Frame into the front and rear returns of the base cabinet If the OPTIONAL HEIGHT FILLER FRAME IS REQUIRED it is installed by aligning it s bottom 3 4 return on the ba...

Страница 11: ...tion that protrudes through the hole in the base cabinet back The cupsink assembly is designed to fit two conditions A When the fume hood assembly is supplied with the Optional Height Filler Frame the...

Страница 12: ...Average B Lighter than Air or high thermal loading C Heavier Than Air Position A is for normal exhaust requirements Locate baffle in center of movement arc All slots are functional This baffle positi...

Страница 13: ...and side panel Lift and lock baffle into position 3 Place lower baffle into proper position Lift and lock baffle into lower support brackets Installing Remote Exterior Control Baffle 1 Place top baffl...

Страница 14: ...holes in track assembly 3 Fasten track with No 8 X 3 8 PPHSMS 4 Locate upper retainer on rear flange of louvered panel Locate retainer 91 29 32 from inside edge of front vertical panel Drill 128 diame...

Страница 15: ...rack with No 8 X 3 8 PPHSMS 4 Locate upper retainer on rear flange of louvered panel Locate retainer 91 29 32 from inside edge of front vertical panel Drill 128 diameter holes using retainer as templa...

Страница 16: ...r 1 Fasten support brackets 1 in front top of frame assembly in holes provided Use 1 4 20 X 3 4 hex head thread cutting screws HHTCS 2 furnished in bag assembly to fasten brackets 2 Fasten rear suppor...

Страница 17: ...1 in front top of frame assembly in holes provided Use 1 4 20 X 3 4 hex head thread cutting screws HHTCS 2 furnished in bag assembly to fasten brackets 2 Fasten rear support 3 to rear top of frame ass...

Страница 18: ...holes provided Use 1 4 20 X 3 4 hex head thread cutting screws HHTCS 2 furnished in bag assembly to fasten brackets 2 Fasten rear support 3 to rear top of frame assembly Use No 8 X 5 8 Phillips pan h...

Страница 19: ...iezoelectric alarm sounds and a red indicator light illuminates to warn of potentially dangerous low air flow conditions 54L0405 Low flow set point Audible and visual alarm I O options 54L0335 Analog...

Страница 20: ...ing from the owner s industrial hygienist or safety representative 2 This fume hood is not intended to be used with all chemicals or all chemical processes Consult the owner s industrial hygienist or...

Страница 21: ...and replacement of broken worn or non functioning parts as needed The following items should be inspected and serviced at least semi annually Liner and baffles for condition and cleanliness Low air f...

Страница 22: ...as shown above lift up and outward to remove post Electrical fixtures are connected to post with flexible conduit and can remain attached CLEANING FUME HOOD INTERIORS Fume hood liners are maintained b...

Страница 23: ...nd of hood is clear See Page 4 Removing front louver panel See Page 21 and 22 Removing blower enclosure panels See Page 16 18 2 Remove the outer portion of the sash moulding strip on all four sides of...

Страница 24: ...at each end 2 Pull bottom out and lift entire panel up to remove 3 Remove both front corner posts See Page 22 WARNING If cable is frayed or damaged it MUST be replaced to avoid personal injury or dama...

Страница 25: ...om frame member and press into place Replace the stainless steel edge protector on the top horizontal edge of the glass Replace the sash guide and remove clamps holding the sash cables Move the sash u...

Страница 26: ...ILIARY AIR FUME HOODS The supply chamber air filter is located in the supply air chamber The change out of the filter requires the removal of the louver panel It is the panel consisting of all louvers...

Страница 27: ...evel Adjust micrometer knob for zero reading as required Add 826 sp gr red gauge oil when needed to maintain zero reading Oil can be obtained from Fisher Hamilton Proper use of the manometer assembly...

Страница 28: ...hoods These tests should be performed by qualified personnel to verify proper operation of the fume hoods before they are put to use The tests of the fume hoods should be performed after the installat...

Страница 29: ...establish an as manufactured and an as used performance rating including the influences of the laboratory arrangement and it s ventilating system The test presumes a conditioned environment No test c...

Страница 30: ...Location ________________________________________ Order Number ____________________________________ Room _______________ Item ______________________ Fume Hood Identification __________________________...

Страница 31: ...ions are offered to help pinpoint and correct the problem Room Cross Drafts Air moving through an open door located adjacent to the fume hood can cause cross drafts An open window or a room air supply...

Страница 32: ...fisherhamilton com DO NOT DISCARD IMPORTANT TEST AND CALIBRATION DATA ENCLOSED TO BE REMOVED ONLY BY HOOD USER PL 326 13 November 14 2003 Part No 55960 Copyright 2003 Fisher Hamilton L L C REMOVE SHIP...