
14
Dimensions are nominal and illustrations and specifications are based
on the latest product information available at the time of publication.
The right is reserved to make changes at any time without notice.
SAFEAIRE
®
LABORATORY FUME HOODS
OPERATION, MAINTENANCE AND INSTALLATION INSTRUCTIONS
INSTALLATION
INSTALLING SAFETY SHIELDS IN RESTRICTED BYPASS, CONSTANT VOLUME BYPASS, AND RADIOISOTOPE
FUME HOODS
1. Locate lower track assembly according to dimensions
shown.
2. Use lower track assembly as template for locating holes.
Drill .128" diameter holes in sill as illustrated, through
existing holes in track assembly.
3. Fasten track with No. 8 X 3/8" PPHSMS.
4. Locate upper retainer on rear flange of louvered panel.
Locate retainer .91" (29/32") from inside edge of front
vertical panel. Drill .128" diameter holes using retainer as
template. Fasten with No. 8 X 3/8" PPHSMS.
5. Position and fasten bumpers using No. 8-32 X 3/4" PPHMS,
washers and hex nuts.
6. Install safety shield by inserting top into upper retainer
channel, then lowering bottom (with rollers) into lower track
assembly.
Louvered Panel
8 X 3/8" PPHSMS
8 X 3/4" PPHSMS
Retainer
Bumper
Safety Shield
8 X 3/8" PPHSMS
Lower Track
Sill
Work Surface
Lower Track
Upper
Retainer
.91"
1.87"
.91"
Hood Width
1.87"
Left Bumper
Right Bumper
.31"
Hood
Lower
Upper
Width
Track
Retainer
48.0"
34.62"
36.56"
60.0"
46.62"
48.56"
72.0"
58.62"
60.56"
96.0"
82.62"
84.56"
1.375"
Содержание SAFEAIRE series
Страница 1: ......