4 | Technical data
FISCHER Mess- und Regeltechnik GmbH
14/24
BA_EN_ME12_ATEX
4 Technical data
4.1 Generalities
Reference conditions (acc. to IEC 61298-1)
Temperature error
+15 … +25 °C
Relative humidity
45 … 75 %
Air pressure
86 … 106 kPa
860 … 1060 mbar
Auxiliary energy
24 V DC
Installation position
User-defined
4.2 Input variables
Messgröße
Pressure in non-aggressive liquid and gaseous media.
Relative pressure
Measuring
Pressure safety
Characteristic
curve devi-
ation
[bar]
[kPa]
Overpressure
Bursting
pressure
Option
0 … +0.6
0 ... 60
4 bar
7 bar
---
0 … +1
0 ... 100
4 bar
7 bar
0.5%FS
0 … +1,6
0 ... 160
4 bar
7 bar
0.5%FS
0 … +2.5
0 ... 250
10 bar
15 bar
0.5%FS
0 … +4
0 ... 400
10 bar
15 bar
0.5%FS
0 … +6
0 ... 600
20 bar
35 bar
0.5%FS
0 … +10
0 ... 1000
40 bar
70 bar
0.5%FS
0 … +16
0 ... 1600
40 bar
70 bar
0.5%FS
0 … +25
0 ... 2500
100 bar
150 bar
---
0 … +40
0 ... 4000
100 bar
150 bar
---
0 … +60
0 ... 6000
200 bar
250 bar
---
Absolutdruck
Measuring
Pressure safety
Characteristic
curve devi-
ation
[bar]
[kPa]
Overpressure
Bursting
pressure
Option
0 … +1
0 ... 100
4 bar
7 bar
0.5%FS
0 … +1.6
0 ... 160
4 bar
7 bar
0.5%FS
0 … +2.5
0 ... 250
10 bar
15 bar
0.5%FS
0 … +4
0 ... 400
10 bar
15 bar
0.5%FS
0 … +6
0 ... 600
10 bar
15 bar
0.5%FS
0 … +10
0 ... 1000
20 bar
35 bar
0.5%FS
0 … +16
0 ... 1600
20 bar
35 bar
0.5%FS