FISCHERSCOPE
®
X-RAY
121
Calibration
Nominal value x measured
s measured min. value
max. value
Evaluation
Cr 1.5
μ
m
1.51
μ
m
0.03
μ
m
1.45
μ
m
1.57
μ
m
OK
Cr 0.77
μ
m
0.75
μ
m
0.02
μ
m
0.68
μ
m
0.79
μ
m
Not OK
The calibration was successful if all re-measurements are within the respective ranges
and you are able to use the measuring application.
If a standard is outside the range, check to see of this measurement range is relevant
for your measuring application. If the measurement range is not relevant for your
measuring application, you can still use the measuring application.
If several re-measurements are outside the range, check the data and potential error
sources. Then perform a new calibration.
For special measuring applications, other conditions may apply for a successful cali-
bration. However, these will be stated separately.
Содержание FISCHERSCOPE X-RAY 4000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Components...
Страница 24: ...24 FISCHERSCOPE X RAY Manual Measurements Deleting Measurement Readings...
Страница 28: ...28 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 44: ...44 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 116: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 116 FISCHERSCOPE X RAY...
Страница 122: ...122 FISCHERSCOPE X RAY Calibration...
Страница 140: ...140 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 166: ...166 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 167: ...WinFTM 167...