FISCHER Mess- und Regeltechnik GmbH
Assembly | 3
BA_EN_DE49_0
11 / 32
2-conductor circuit
Input signal
Non-ATEX
supply
+Sig
-Sig
+U
b
-U
b
Output signal
Pressure transmitter
DE49
+
Intrinsically safe
barrier
+
Δp
ATEX zone
ia
Illustration 6:
2L Circuitry
Supply and signal power circuit limit values
(Ignition protection type intrinsic safety Ex ia IIC)
U
i
≤ 30 V
I
i
≤ 100 mA
P
i
≤ 750 mW
inner effective capacity
C
i
2.5 nF
effective inner inductivity
L
i
negligible
NOTICE! In contrast to the EMV-GND, the power connections have an inner
capacity of max. 5 nF.
Pin assignment
Illustration 7:
M12 connector 5-pin
Pin
Signal name
Cable colour
1
Supply (+) / output (+)
+U
b
+Sig
brown
2
unused
n.c.
white
3
Supply (-) / output (-)
-U
b
-Sig
blau
4
unused
n.c.
schwarz
5
Functional earth
green/yellow
A
Coding A
Table 1:
Supply and output signal