7.
CAMERA DESCRIPTION AND START UP
Please refer to the following figure and follow the order listed below, before connecting your camera:
Cooling
The camera can be air cooled or liquid cooled (no liquid nitrogen) (1).
By default, the camera activates its fan.
Air cooling allows to cool the detector down to nominal -15°C with a maximum room temperature of 35°C.
Air cooling is ensured by the fan, and the fan speed can be controlled in automatic mode or manual mode.
Liquid cooling (1) allows to cool the detector down to nominal -40°C with the following conditions.
Cooling System
Parameters
Value
Flow rate
≥ 0.5 L/min
Pressure
10 bar max
Liquid Temperature
35°C max
Approved Fluids
Ethylen Glycol aqueous solution (Max concentration: 50%)
Distilled water or deionized water (Min temperature 10°C)
Cooling capacity
100 W minimum
Heat is evacuated by circulating a cooling fluid through two ports (G1/8 thread). To fasten your connectors,
First Light Imaging recommends using Loctite 577 thread sealant.
The cooling hoses can be fastened in any order.
After turning on the cooling unit, please check that no leaks are visible.