7.

 

CAMERA DESCRIPTION AND START UP 

 

Please refer to the following figure and follow the order listed below, before connecting your camera:  
 

 

 
 

 

Cooling  

 

The camera can be air cooled or liquid cooled (no liquid nitrogen) (1).   

By default, the camera activates its fan.  
 
Air cooling allows to cool the detector down to nominal -15°C with a maximum room temperature of 35°C. 
Air cooling is ensured by the fan, and the fan speed can be controlled in automatic mode or manual mode.  

 

Liquid cooling (1) allows to cool the detector down to nominal -40°C with the following conditions.  

 

Cooling System 
Parameters 

Value 

Flow rate 

≥ 0.5 L/min  

Pressure 

10 bar max 

Liquid Temperature 

35°C max 

Approved Fluids 

Ethylen Glycol aqueous solution (Max concentration: 50%) 
Distilled water or deionized water (Min temperature 10°C) 

Cooling capacity  

100 W minimum 

 

Heat is evacuated by circulating a cooling fluid through two ports (G1/8 thread). To fasten your connectors, 
First Light Imaging recommends using Loctite 577 thread sealant. 
The cooling hoses can be fastened in any order. 
 

After turning on the cooling unit, please check that no leaks are visible.

 

 

Содержание C-RED 2

Страница 1: ...C RED 2 Quick Start Manual C RED 2 Quick Start_20180622...

Страница 2: ......

Страница 3: ...the instructions of use 2 SYMBOLS AND INDICATIONS Please read this Quick Start guide and the following definitions carefully to understand the potential dangers and the precautions to take Please refe...

Страница 4: ...the system assembler only For your safety the equipment must be TURNED OFF AND UNPLUGGED before any technical intervention The security provided with this equipment is only guaranteed with a use in a...

Страница 5: ...rent accessory packs Items may differ from pictures Item name Quantity Picture Camera 1 Power supply 1 IEC cable 1 USB C to USB A cable 1 USB key with User manual Demo software Test report 1 Quick sta...

Страница 6: ...l mode Liquid cooling 1 allows to cool the detector down to nominal 40 C with the following conditions Cooling System Parameters Value Flow rate 0 5 L min Pressure 10 bar max Liquid Temperature 35 C m...

Страница 7: ...ted with specific grabbers that we highly recommend using Please refer to the demo software manual List of tested recommended grabbers MATROX Radient eV cl full Teledyne DALSA X64 Xcelera CL Windows o...

Страница 8: ...cription Configuring Blue Camera starting Operational Green Camera configuration is applied Operational cooling White Camera is cooling down Operational cold Purple Camera has reached target temperatu...

Страница 9: ...adout function The acquisition speed can be set to any value from 0 001 to 600 fps The integration time can also be set The integration time range is 50 s 1 fps For integration time below 50 s please...

Страница 10: ...e protection on all power lines Any connected equipment should be powered off before removing any connection between the computer and C RED 2 14 MAINTENANCE Cleaning of window Never use an unclean clo...

Страница 11: ...2138 USA Tel 33 4 42 61 29 20 E mail support first light fr Website www first light us For the rest of the world FIRST LIGHT IMAGING SAS Europarc Sainte Victoire B t 6 Route de Valbrillant Le Canet 13...

Страница 12: ...Notes...

Страница 13: ...Notes...

Страница 14: ......

Отзывы: