Before your OEM vi30/40 laser is put into service for the first time, its functionality 

should be verified. 

Proceed to the next section or the following procedure if you are not utilizing the UC-

2000 controller to verify the laser system is operating at optimum performance. 

Important 

Note:

  

Controlling your laser without a UC-2000 controller:

Caution 

Possible 

Equipment 

Damage 

Do not reverse polarity when connecting DC power cables to your DC power 

source. 

Reversed DC polarity will damage the lasers internal RF and control board 

circuitry. Carefully follow the directions below to ensure that DC cable leads are 

properly connected to the correct DC output terminals.

Turn off DC power before installing or removing any plug or cable

 

from the DB-9 

I/O connector.

 Ensure that user connections are made to the appropriate pins and 

that the appropriate signal levels are applied. Failure to do so may damage the 

laser.

Vi-Series lasers MUST be provided with a pre-ionizing “tickle” signal 

during 

standby or laser “low” periods. This signal is automatically provided by SYNRAD 

UC-2000 Universal Laser Controller.

A tickle signal keeps the plasma ionized during laser ‘low’

 periods and facili-

tates plasma breakdown and pulse-to-pulse fidelity. Damage to or malfunction of 

the laser may occur if this, or equivalent drive signals are not provided. 
If the laser has been 

“OFF” or the tickle has not been applied for any length 

of time 

apply the tickle signal for at least two seconds before applying any PWM 

command signal.

Warning!

Serious Personal 

Injury

Always, use lockout/tagout for safety. Always wear safety goggles. Always 

use shielded cable when connecting your PWM Command signal source to 

PWM Positive/PWM Negative inputs.

 In electrically-noisy environments, long lengths of unshielded wire act like 

an antenna and may generate enough voltage to trigger uncommanded 

lasing.

Control Connections (Continued):

Содержание vi30

Страница 1: ...ossible then immediately notify the shipping carrier responsible party and SYNRAD After unpacking review the Inventory list in the Intro section within the operators manual If you need a replacement b...

Страница 2: ...one metric or SAE fastener per mounting tab on the baseplate Do not use any type of jackscrew arrangement as this will twist the baseplate and may distort the tube Important Note SYNRAD does not reco...

Страница 3: ...v30 height for upgrading from v30 to vi30 40 for mounting from above The Tall Wide mounting kit SYNRAD P N 250 20190 02 6 Standard Mounting Drill three thru holes in the final mounting surface that co...

Страница 4: ...unting Feet Matches v30 Beam Exit Height Tall Wide Mounting Feet Matches v30 Beam Exit Mounting When mounting the Firestar laser use only one metric or SAE fastener per mounting tab on the baseplate D...

Страница 5: ...t be less than 3 0 in 76 mm Minimum fan to wall clearance is 2 25 in 57 2 mm otherwise damage will occur 3 0 in 76 mm 3 0 in 76 mm Front 8 If temperature monitoring is desired see the following figure...

Страница 6: ...he Technical Reference section of the laser s Operation Manual Airflow Direction Front to Rear Allow 6 inches clearance Recommended fan shroud design for rear cooling Caution Possible Equipment Damage...

Страница 7: ...know Our lasers incorporate the following wetted materi als in the coolant path nickel plated brass copper Acetal nitrile Viton PBT Delrin Polyethelene and stainless steel Avoid glycol based additive...

Страница 8: ...no more than 6 6 mm 0 26 in into the cooling manifold Maintain coolant pH above 7 Operating the laser with a coolant temperature below the dew point of the surrounding air may cause condensation to o...

Страница 9: ...ce the humidity of the enclosure containing the laser D Install gas purge Introduce nitrogen or dry filtered air into the laser housing See Assist gas purity and Purge gas specifications in the Techni...

Страница 10: ...the laser chassis serves as DC power ground If you need to isolate the body of the laser from the equipment chassis ground use an insulating pad or film and insulated mounting hardware OEM vi30 40 la...

Страница 11: ...the laser s safety interlock function potentially exposing personnel to hazardous invisible laser radiation If using a UC 2000 Universal Laser Controller refer to the UC 2000 Laser Controller Quick S...

Страница 12: ...to the appropriate pins and that the appropriate signal levels are applied Failure to do so may damage the laser Vi Series lasers MUST be provided with a pre ionizing tickle signal during standby or...

Страница 13: ...ated PWM Command signal to the laser s CTRL input connector Refer to Controlling laser power in the Technical Reference chap ter in your laser s Operation Manual for tickle and PWM Command signal desc...

Страница 14: ...uency of 5 kHz at a 5 volt level and ensure that the controller s duty cycle is set to zero percent output 0 0 23 Apply a Laser Enable signal to the DB 9 I O connector The yellow RDY lamp turns on to...

Отзывы: