The FC-IR101 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the FC-IR101 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the FC-IR101 as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
Separation distance according to frequency of transmitter
m
Rated
maximum
output power
150kHz to 80MHz
d
=1.2
P
80MHz to 800MHz
d
=1.2
P
800MHz to
2.5GHz
d
=2.3
P
0.01
0.01
0.12
0.23
0.1
0.1
0.38
0.73
1
1
1.2
2.3
10
10
3.8
7.3
100
100
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
14. Commitment to quality and after-sales service
This product provides 1 year free maintenance service.
Note: The free service does not include the failure and damage due to user’s personal
reasons or unauthorized disassembling.
Caution
Please keep the purchase vouchers to facilitate future maintenance.
15. Manufacturer information
Shenzhen Finicare Co., Ltd
Address: Room 201, A Building, NO.1 Qianwan Road, Shengang Cooperative Zone, Qianhai,
Shenzhen, China
Tel: +86-755-23013503
Fax: +86-755-23574425
Version: 1.0
Issue date: 02/01/2018