1.
Switch off the monitor and disconnect it from the power line.
2.
Wipe the entire exterior surface, including the handle, battery cover and compartment,
LCD, keys and inside surface of the protective cover, of the equipment using a soft cloth
dampened with the cleaning solution thoroughly until no visible contaminants remain.
3.
Wipe off the cleaning solution with a fresh cloth or towel, dampened with tap water
after cleaning until no visible cleaning agent remains.
4.
Dry the monitor in a ventilated and cool place.
11.3 Disinfection
If the point(s) has been in contact with the patient without the replaceable ear cover
co-packaged, e.g. surface of infrared probe and protective cover, then cleaning and
disinfection is required after every use to avoid patient cross infection. If there has been no
patient contact and there is no visible contamination, including the handle, battery cover
and compartment, LCD, keys and inside surface of the protective cover, or when the probe
is used with the replaceable ear cover, daily cleaning and disinfection is appropriate, and
the suggestion is do it weekly.
For the whole unit, low level disinfection is appropriate since it is a complete surface
contacting device. Clean the device before disinfection. The validated disinfectant for
cleaning the device is:
Isopropanol (70%)
Since Isopropanol is used for both cleaning and disinfecting, a new cloth is required to be
used for the disinfection step.
To disinfect the monitor, follow these steps:
1.
Switch off the monitor and disconnect it from the power line.
2.
Wipe the display screen using a soft, clean cloth dampened with the disinfectant
solution.
3.
Wipe the exterior surface of the equipment using a soft cloth dampened with the
disinfectant solution.
4.
Wipe off the disinfectant solution with a dry cloth after disinfection if necessary
5.
Dry the monitor for at least 30 minutes in a ventilated and cool place.
12. Troubleshooting
Table 4
Diagnosis
Solution
The screen shows " Hi " or
"Lo”
1. Check the measurement object. It is unable to ensure
the measurement of the forehead hair, water,
sweating, and applying cosmetics case.
2. Check the operation environment. Environment
changes will greatly influence the measurement.
3. Temperature change is too big or the thermometer
tests target of low temperature immediately after
being switched from a high temperature measuring.
It's better to use after 5 minutes to achieve a new
heat balance.