background image

Revised 10/12

4

TECHNICAL INFORMATION FOR CDM4160

is pushed during warm-up time (within the first two 

hours after the module is powered on), the baseline 

voltage is read at that moment and the module 

immediately goes into operation mode.

Caution:

 If the baseline reset switch is pushed while 

in a polluted environment where the actual CO

2

 

concentration is higher than 400 ppm (ambient 

levels), the accuracy of readings may become 

adversely affected.

2-7  Input-output signal (CN1)

Please refer to Table 4-Rated input and output voltages 

for rated values. Also please refer to Fig. 2-Circuit 

diagram of interface

 for circuit.

2-7-1  Vin (Pin No. 1)

Regulated voltage should be input into this port. The 

sensor’s output may vary according to the sensor’s 

voltage dependency characteristics if the input 

voltage fluctuates. Please refer to "TGS4160 Technical 

Information"

 for details.

2-7-2  Concentration output (Pin No. 2)

An analog voltage corresponding to CO

2

 concentration 

is output from this port according to the following 

equation:

 

CDM4160-L00 : Vconc=CO

2

 conc/1000

 

CDM4160-M00 : Vconc=CO

2

 conc/2000

 

CDM4160-H00 : Vconc=CO

2

 conc/10000 

 

2-7-3  Control signal output (Pin No. 3)

When CO

2

 concentrations exceed the control 

concentration, output from this port will be “ON”.

Table 4 - Rated input/output voltages

Table 3 - CDM4160 jumper pin settings

2-7-4  Trouble signal output (Pin No. 4)

This port will output “ON” in the following 

situations:

 

*  when sensor output is abnormal

 

*  when the connection for control concentration   

              setting is broken

 

2-8  LEDs (refer to Table 5)

2-8-1  Green LED (LED 1)

The green LED indicates the module’s power 

condition. It will be lit when the power is on. The 

LED blinks on and off during warm-up time (for the 

first two hours after power is turned on).

2-8-2  Yellow LED (LED 2)

The yellow LED indicates a trouble condition. It blinks 

on and off when trouble signal output is “ON”.

2-8-3  Red LED (LED 3)

The red LED indicates that CO

2

 concentration is over 

the control concentration level. It lights when control 

signal output is “ON”.

Pin 

No.

Item

Minimum

Typical

Maximum

u/m

1

Power supply input (V

IN

)

3.5

5.0

5.5

V

2

CO

2

 conc. 

output 

(V

CONC

)

Output voltage

GND

-

4.6

V

3

Control signal 

output

(CTRL)

Output voltage 

(trouble)

I

OL

=1mA

GND

-

0.2

V

Output voltage 

(warm-up) I

OH

=1mA

V

IN

-0.6

-

V

IN

V

Allowable current

-

-

25

mA

4

Trouble signal

output

(TRBL)

Output voltage

I

C

=5mA

-

-

0.3

V

Allowable voltage

-

-

50

V

Allowable current

-

-

100

mA

Mode

JP1

JP2

Threshold CO

2

 Concentration of Control Signal

CDM4160-L00

CDM4160-M00

CDM4160-H00

ON

OFF

ON

OF

ON

OFF

I

OPEN

OPEN

800ppm

720ppm

1,000ppm

900ppm

5,000ppm

4,500ppm

II

OPEN

SHORT

1,000ppm

900ppm

2,000ppm

1,800ppm

10,000ppm

9,000ppm

III

SHORT

OPEN

1,500ppm 1,350ppm

5,000ppm

4,500ppm

20,000ppm

20,000ppm

IV

SHORT

SHORT

2,000ppm 1,800ppm

8,000ppm

7,200ppm

40,000ppm

36,000ppm