SPECIFICATIONS
Display:
3½ digit liquid crystal display (LCD) with a
maximum reading of 1999.
Overrange:
"OL" mark indication.
Auto power off:
60 minutes.
Operating environment:
32 to 122
°
F (0 to 50°C) at
<70% R.H.
Storage temperature:
-4 to 140
°
F (-20 to 60°C), 0
to 80% R.H. with battery removed.
Accuracy:
Specifications good in ambient condi-
tions of 73
°
F
±
9
°
F (23°C
±
5°C), <75% relative
humidity.
Temperature Coefficient:
0.1
×
(specified accuracy)
per
°
F/°C. (32 to 64
°
F (0 to 18°C), 82 to 122
°
F
(28 to 50°C)).
Power:
Single standard 9-volt battery, NEDA 1604,
JIS 006P, IEC 6F22.
Battery life:
300 hours typical with alkaline.
Accessories:
One pair test leads, one pair alligator
clips, k-type thermocouple (HS33), 9V battery
(installed), and operating instructions.
Safety:
UL, CE, Cat III 600V, UL3111,
IEC/EN61010-1, C-Tick certified.
Diode test
Test current:
∼
1.0mA
Accuracy:
±
(1.5% rdg + 3 dgts)
Open circuit volts:
3.0VDC typical
Overload protection:
500VDC or AC rms
Continuity
Audible indication:
Less than 100
Ω
Response time:
100ms
Green LED will be on continuously.
Overload protection:
500VDC or AC rms
Capacitance
Range:
200
μ
F
Resolution:
0.1
μ
F
Accuracy:
±
(3% rdg + 5 dgts)
Overload protection:
500VDC or AC rms
DC volts
Ranges:
200mV, 2000mV, 20V(HS32), 200V
Resolution:
0.1mV
Accuracy:
±
(0.5% rdg + 1 dgt)
Input impedance:
560k
Ω
on V inputs, 10M
Ω
on
mV input
Overload protection:
600VDC or AC rms,
500VDC/350VAC rms 15 sec on 200mV range
Transient protection:
6kV for 10
μ
sec
AC volts (50Hz - 500Hz)
Ranges:
200mV, 200V, 600V
Resolution:
0.1mV
Accuracy:
±
(1.2% rdg + 3 dgts)
±
(2.0% rdg + 5
dgts) on 600V range
Input impedance:
560k
Ω
on V input, 10M
Ω
on mV
input
Overload protection:
600V DC or AC rms
Transient protection:
6kV for 10
μ
sec
Resistance
Ranges:
200
Ω
, 200k
Ω
Resolution:
0.1
Ω
Accuracy:
±
(1.0% rdg + 4 dgts)
Open circuit volts:
0.3VDC typical, (3.0VDC on
200
Ω
range)
Overload protection:
500VDC or AC rms
Temperature (HS33)
Range:
-30 to 200
°
F (-34 to 93°C)
Resolution:
0.1
°
F/°C
Accuracy:
±
1
°
F, 32 to 120
°
F (0 to 48°C),
±
1% + 1.5
°
F, -4 to 200
°
F (-20 to 93°C),
±
2% + 3
°
F, -30 to -4
°
F (-34 to -20°C).
Sensor type:
K-type thermocouple
Overload protection:
60 VDC or 30 VAC rms
Available Fieldpiece access. heads
There is a Fieldpiece accessory head available for
just about any job. There are heads to measure tem-
perature, RH%, wet bulb, dew point, vacuum
(microns of mercury), manometer (inches of water
column), amps AC&DC, high voltage, CO, CO2, air
velocity, and many more.
Using & storing test leads
Because the wire insulation is silicone the leads
will stay flexible in cold weather and will not melt if
bumped by a soldering iron.
Disconnect top half of test lead and plug tip direct-
ly into meter to make voltage testing easy. Use with
included alligator clip (ASA2) as shown for even eas-
ier operation.
For convenient lead storage, wrap the leads as
shown. Pull leads around front between overhanging
tips, twist, and pull over one of the lead plugs.
Voltage
Capacitance
Resistance
Selecting Ranges
For DC voltage, set the meter to the VDC parame-
ter instead of VAC as shown above.
For all ranges choose a range just above the value
you expect. If display reads “OL” (overload), select a
higher range. If display shows less than three num-
bers, select a lower range for better resolution.
Temperature (HS33)
Works with Fieldpiece accessory
heads!
Connect to Fieldpiece accessory heads by simply
attaching them to the top of meter
(1)
or attach
remotely through leads
(2)
. For most heads, move
dial to range shown
(1)
. For the AAC clamp (ACH4),
move dial to VAC range
(2)
.
Disconnect test
leads from volt-
age before plug-
ging in thermo-
couple!
Insure the tem-
perature being
measured is stable.
Maintain good con-
tact between the
thermocouple and
what's being meas-
ured.
CAT.III
300V
400A
CLAMP
ACH4
AC Current
Clamp
1AAC / 1mVAC
400AAC MAX
!
(1)
(2)
The velcro strip supplied is
intended to be used as shown
on the right. Tie thermocouple
like this and wrap velcro
around surface to make sure
bare wires are in close con-
tact with surface being meas-
ured. This is especially help-
full when taking pipe temps.
The k-type thermocouple
supplied is designed with a
“wrap tab” to keep the wire
out of the way when not
needed. Just wrap the wire
around tab and pull through
the hole at the top to keep in
place as shown to the left. No
more clutter when stored.