Operating Environment:

The controller and accessories are designed to be used in the normal operating room environment and require no special 
handling or care exceptional to other electronic devices used in that environment. The controller and hand held remote control 
should be positioned within 3 meters of but outside the sterile field, the motor unit is sleeved inside a sterile drape sleeve and 
mounted on the drive within the sterile field.
Temperature Range: +5°C to +40°C
Relative Humidity Range:  10% to 95% (non-condensing)
Atmospheric Pressure Range:  500 hPa to 1060 hPa
Altitude: ≤ 2000 m above sea level

 

Storage and Transport Conditions

The microTargeting™ Controller Power Assist System should be stored and transported within the carrying case provided when 
not in use.

Warnings and Cautions:

WARNING:  To avoid risk of electric shock, this equipment must only be connected to a supply mains power outlet with  
Protective Earth (3 pronged socket), never use a damaged power cord or power strip.

CAUTION:  High voltage - there are no user serviceable parts internal to the controller housing, do not attempt to disassemble 
the controller or any of its accessories.

WARNING:  No unauthorized modification of this equipment is allowed.

WARNING:  Do not block the vents on the bottom or back of the controller as this could cause it to overheat.  Do not remove  
the 4 rubber feet as this will cause the vents on the bottom of the controller to become blocked.

WARNING:  Carefully route all system cabling away from high traffic areas.

WARNING:  Do not manually rotate the drive knobs when the motor is attached as this could potentially damage the motor unit.
Note:  To ensure optimal performance and minimize motor noise, the controller system should be serviced annually to tune the 
controller’s drive circuit and speed settings to compensate for normal motor wear.

WARNING:  Do not use the controller in the presence of flammable gas mixtures.

WARNING:  Do not attempt to sterilize the motor unit or the hand held remote control.

WARNING:  Expected target depth can change based on the stereotaxy and electrodes used.  Ensure the target depth is set  
correctly when running the controller in distance to target mode.

WARNING:  Always zero the controller prior to inserting electrodes.

WARNING:  Periodically (every 5mm is recommended) verify that the depth reported by the controller matches that shown on 
the drive scale.

WARNING:  If power to the controller is temporarily  interrupted, it will need to be re-zeroed prior to resuming its use.

CAUTION:  Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into  
service according to the EMC information provided in the Declaration of Electromagnetic Emissions section of this document.

CAUTION:  Portable and mobile RF communications equipment can affect medical electrical equipment.

WARNING:  The controller should not be used adjacent to or stacked with other equipment.  Where such an arrangement is 
necessary, the controller should be observed to verify normal operation prior to use.

CAUTION:  Federal law (USA) restricts this device to sale by or on the order of a physician.

Rx only

Содержание microTargeting 66-DA-ME

Страница 1: ...ontroller Power Assist System FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 24 hour technical service 1 800 326 2905 US Can 1 207 666 8190 FHC Europe TERMOBIT PROD srl 42A Barbu Vac...

Страница 2: ...Page 2 of 17 L011 80 Rev C0 2020 11 05...

Страница 3: ...Illustrative Procedure Basic pre use checkout Prepare the controller for use with the drive Assemble and connect the motor unit Zero the drive Typical operative use After completion of the procedure S...

Страница 4: ...It interfaces with the micropositioner responsible for positioning a microelectrode in the brain WARNING Caution consult instructions for important cautionary information Medical device manufacturer...

Страница 5: ...Carefully route all system cabling away from high traffic areas WARNING Do not manually rotate the drive knobs when the motor is attached as this could potentially damage the motor unit Note To ensure...

Страница 6: ...5 Slide fuse drawer with fuses into power entry Spring tabs will snap into place once drawer is fully seated Installation and Functional Checkout Prior to its initial use set up the microTargeting Co...

Страница 7: ...ide of the drape 5 Push the drape over hand so the motor and cable are all the way at the end of the sleeve 6 and Maneuver the drape and motor so that the two alignment pins and the center drive plate...

Страница 8: ...and are securely attached to the cables they terminate Position the controller module such that the mains power switch on the back panel is easily accessible Prepare the controller for use with the dr...

Страница 9: ...bly any farther at this time 8 Push down lightly on the motor unit while turning the drive advancement knob slowly 9 The motor unit s center drive plate pins must be exactly aligned with the mating ho...

Страница 10: ...e Set drive origin message 14 If an initial offset is to be used it should be programmed into the controller now Note that the initial offset will reset to 0mm every time the power to the controller i...

Страница 11: ...d motion speed using the speed selection slider When set to 0 no motion will occur Sensible defaults are provided for the other speed steps but they can be adjusted through the USB Interface Pressing...

Страница 12: ...r unit and all its accessories in the carrying case Stall Detection The torque of the stepper motor is quite high especially when amplified by the drive screw but the motor can be made to stall if it...

Страница 13: ...artment before attempting any repairs or alterations Many of these repairs will still be performed at the factory at no charge to the customer Service Repairs Should service be required please contact...

Страница 14: ...ation Sterilization AAMI IEC 60601 3rd Ed Do not sterilize Performance Specifications Linear Resolution Long Term Linear Accuracy Minimum Speed Maximum Speed Acceleration Deceleration Travel Rates Saf...

Страница 15: ...l offset target and limit depths will be displayed by the controller when the zero button is pressed any time after zeroing the drive Speeds There are four speed options available they can be set from...

Страница 16: ...thetic the r h should be at least 30 EFT IEC 61000 4 4 2kV Mains 1kV I Os 2kV Mains 1kV I Os Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV...

Страница 17: ...distance between portable and mobile RF Communications Equipment and the mT Controller as recommended below according to the maximum output power of the communications equipment Max Output Power Watts...

Отзывы: