12.  Plug the motor unit into the controller by inserting its connector into the corresponding socket on the front 

panel. 
 
 

 
 
 
 
 
 
 
 
 
Zero the drive

13.  Activate the ON/OFF switch of the controller module. The controller should boot and display the “Set drive origin” 

message.  
 
 
 
 
 
 
 
 
 
 
 

14.  If an initial offset is to be used, it should be programmed into the controller now.  Note that the initial offset will 

reset to 0mm every time the power to the controller is cycled.  

15.  Select a speed using the speed selection slider on the remote control. Using the rocker switch, advance or retract 

the drive as needed until the drive is positioned at exactly 0 µm or the initial offset location if specified.

Rocker switch

Speed selection slider

 

Содержание microTargeting 66-DA-ME

Страница 1: ...ontroller Power Assist System FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 24 hour technical service 1 800 326 2905 US Can 1 207 666 8190 FHC Europe TERMOBIT PROD srl 42A Barbu Vac...

Страница 2: ...Page 2 of 17 L011 80 Rev C0 2020 11 05...

Страница 3: ...Illustrative Procedure Basic pre use checkout Prepare the controller for use with the drive Assemble and connect the motor unit Zero the drive Typical operative use After completion of the procedure S...

Страница 4: ...It interfaces with the micropositioner responsible for positioning a microelectrode in the brain WARNING Caution consult instructions for important cautionary information Medical device manufacturer...

Страница 5: ...Carefully route all system cabling away from high traffic areas WARNING Do not manually rotate the drive knobs when the motor is attached as this could potentially damage the motor unit Note To ensure...

Страница 6: ...5 Slide fuse drawer with fuses into power entry Spring tabs will snap into place once drawer is fully seated Installation and Functional Checkout Prior to its initial use set up the microTargeting Co...

Страница 7: ...ide of the drape 5 Push the drape over hand so the motor and cable are all the way at the end of the sleeve 6 and Maneuver the drape and motor so that the two alignment pins and the center drive plate...

Страница 8: ...and are securely attached to the cables they terminate Position the controller module such that the mains power switch on the back panel is easily accessible Prepare the controller for use with the dr...

Страница 9: ...bly any farther at this time 8 Push down lightly on the motor unit while turning the drive advancement knob slowly 9 The motor unit s center drive plate pins must be exactly aligned with the mating ho...

Страница 10: ...e Set drive origin message 14 If an initial offset is to be used it should be programmed into the controller now Note that the initial offset will reset to 0mm every time the power to the controller i...

Страница 11: ...d motion speed using the speed selection slider When set to 0 no motion will occur Sensible defaults are provided for the other speed steps but they can be adjusted through the USB Interface Pressing...

Страница 12: ...r unit and all its accessories in the carrying case Stall Detection The torque of the stepper motor is quite high especially when amplified by the drive screw but the motor can be made to stall if it...

Страница 13: ...artment before attempting any repairs or alterations Many of these repairs will still be performed at the factory at no charge to the customer Service Repairs Should service be required please contact...

Страница 14: ...ation Sterilization AAMI IEC 60601 3rd Ed Do not sterilize Performance Specifications Linear Resolution Long Term Linear Accuracy Minimum Speed Maximum Speed Acceleration Deceleration Travel Rates Saf...

Страница 15: ...l offset target and limit depths will be displayed by the controller when the zero button is pressed any time after zeroing the drive Speeds There are four speed options available they can be set from...

Страница 16: ...thetic the r h should be at least 30 EFT IEC 61000 4 4 2kV Mains 1kV I Os 2kV Mains 1kV I Os Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV...

Страница 17: ...distance between portable and mobile RF Communications Equipment and the mT Controller as recommended below according to the maximum output power of the communications equipment Max Output Power Watts...

Отзывы: