9.  Select “Microstim” in the dialog box.
10.  Select the appropriate channel in the dialog box.  Verify that the test electrode is still plugged into this channel.
11.  Set the stimulation amplitude to 100 µA. Set the pulse duration to 1000 µsec. Set the stimulus duration to fixed duration  
 

and select 5 sec. Press the orange activate button on the remote control. Confirm that the software displays the warning  

 

“Compliance Voltage Reached.” Confirm that the stimulus monitor tone is audible and set to a comfortable volume.

12.  Switch to “Macrostim” mode.
13.  Set the stimulation amplitude to 5 mA.  Set the pulse duration to 1000 µsec.  Press and hold the orange activate button on  
 

the remote control to start stimulation.

14.  Confirm that the current output is 5 ± 0.2 mA, the voltage output is 5 ± 0.5 V, and that stimulation stops when the orange  
 

button is released.  

15.  Close the stimulation dialog box.
16.  Verify the impedance check and stimulation functionality of each  available channel (by repeating step 3B) before every  
 surgery.

WARNING: If an error occurs during the pre-use checkout procedure (step 3), discontinue use of the system. Contact an FHC 
factory authorized representative for further instructions.

WARNING: Failure to properly perform the pre-use checkout procedure may result in patient injury or death.

WARNING: Disconnect the test electrode before connecting the LP+ to a patient.

4.  Load a user profile and enter patient information.

a.  Remove the test electrode if it is still attached to the LP+.
b.  Click on the “Patient” tab at the top of the screen.  Select a profile at the bottom of the screen.  Click “Apply Profile” to load the  
 profile.
c.  Click “Add New Patient” on the left-hand side of the screen to create a new patient.
d.  Enter patient information in the fields provided.

5.  Enter equipment and targeting information.

a.  Click on the “Setup” tab at the top of the screen.
b.  Enter equipment information and patient targeting coordinates using the fields provided.

WARNING: Incorrect equipment specifications could lead to errors in targeting or electrode depth.
WARNING: Using a non-supported drive, frame, or other device may result in incorrect targeting.
WARNING: Incorrect zeroing of the microTargeting Controller can cause serious injury or death.
WARNING: Incorrect specification of the initial target value can cause serious injury or death.

6.  Perform microelectrode recording.

A.  Connect electrode cables as shown in the diagram on page two.  Connect electrodes and cables one at a time to avoid  

 confusion.

1.  Assign an electrode to a channel using the Guideline 4000 LP+ software.
2.  Check for a green light next to the channel on the Guideline 4000 LP+.
3.  Connect the sterile electrode cable to the microelectrode, pass the DIN connector end out of the sterile field, and connect  
 

the DIN connector to the correct channel.

4.  Click on the “Electrode” tab at the top of the screen.

Содержание 66-EL-LP

Страница 1: ...products MT LPP MT LPP APM 66 EL LP FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 www fh co com FHC Europe TERMOBIT PROD srl 42A Barbu Vacarescu Str 3rd Fl Bucharest 020281 Sector 2...

Страница 2: ...L011 57 Rev D1 2018 11 09 Page 2 of 11...

Страница 3: ...y Warnings and Cautions System Maintenance and Repair Handling Considerations Maintenance Repair System Disposal System Specifications Performance Accuracy Microelectrode Recording Stimulation Impedan...

Страница 4: ...complies with MDD Directive 93 42 EEC and legal responsibilities as a manufacturer are with FHC Inc 1201 Main Street Bowdoin ME 04287 USA Temperature limitation Storage conditions are between 0 C 40 C...

Страница 5: ...osal Return the Guideline 4000 LP to FHC for proper end of life disposal Please contact an FHC factory authorized representative to arrange a return System Specifications Performance Accuracy Microele...

Страница 6: ...patient isolated AC DC Coupled Bandwidth 0 5kHz Ranges factory configurable 0 to 16V 8 0V to 8 0V 0 to 8 0V 4 0V to 4 0V 0 to 1 6V 800mV to 800mV 0 to 800mV 400mV to 400mV 0 to 160mV or 80mV to 80mV...

Страница 7: ...speakers or Guideline 4000 LP and the patient simultaneously If using headphones use wireless headphones if contact with patient is possible WARNING The Guideline 4000 LP Multiple Portable Socket Out...

Страница 8: ...nfirm that a scrolling waveform is visible in the corresponding thumbnail window for each installed channel a If any channels have failed to initialize properly wait one minute for the LP to detect an...

Страница 9: ...nect the test electrode before connecting the LP to a patient 4 Load a user profile and enter patient information a Remove the test electrode if it is still attached to the LP b Click on the Patient t...

Страница 10: ...n using the orange activate button on the stimulation remote control If desired stimulation may be stopped by pressing the orange activate button again during fixed duration mode only WARNING Electrod...

Страница 11: ...e top center of the screen to view root mean square RMS voltage versus depth and mean firing rate MFR versus depth plots for all channels d Click on the Guidance tab at the top left of the computer sc...

Отзывы: