2.  Power up the Guideline 4000 LP+.

a.  Turn on the computer and log on as “Guideline 4000” using the password provided.
b.  Double click the Guideline 4000 icon on the Windows computer desktop to start the software.
c.  Once the Guideline 4000 opening screen appears, turn on the LP+, speakers, and, if desired, the microTargeting Controller.

WARNING: Discontinue use of the Guideline 4000 LP+ should any erratic function or damage become evident.

CAUTION: Computer systems are periodically unstable. In the event of a system crash, the Guideline 4000 LP+ can be rebooted.  

CAUTION:  Please read the manual accompanying the notebook computer for complete information on the notebook’s use and 
specifications.

WARNING: Do not enable the wireless capabilities of the notebook computer in the hospital environment unless you are authorized 
to do so.

WARNING: Do not install third party software onto the Guideline 4000 LP+ notebook computer unless authorized to do so by an FHC 
factory authorized representative. 

WARNING: Use only the supplied notebook computer to operate the Guideline 4000 LP+.

CAUTION:  The Guideline 4000 LP+ notebook computer comes with pre-installed anti-virus software. Do not perform an anti-virus 
scan before or during a surgical procedure. Should a virus be found on the notebook computer, contact FHC for quarantine and 
removal instructions.

3.  Perform the pre-use checkout procedure.

A.  Verify recording functionality.

1.  Confirm that the channel status LED indicator lights for all available channels on the LP+ are green. The light will only turn 
 

green for installed channels.

2.  Click on the “Electrode” tab on the computer screen.  Confirm that a scrolling waveform is visible in the corresponding 
 

thumbnail window for each installed channel.
a.  If any channels have failed to initialize properly, wait one minute for the LP+ to detect and automatically reinitialize 

   them.

b.  If after one minute a channel has still not initialized, right click on the corresponding thumbnail and select “Re-initialize.”
c.  If a channel has still not initialized, shut down the Guideline 4000 LP+ by selecting  “Shutdown” from the system 

 

 

menu. Once the computer has shut down, turn off the LP+, speakers, and microTargeting Controller.  Restart the 

 

 

system by following steps 1, 2, and 3 above.

3.  Verify that sound is audible for each channel. 

a.  Toggle between muted and unmuted states for the channel using the corresponding numeric key (for example, “1” for 

 

 

channel 1).  The speaker icon above the thumbnail will indicate the state of the channel.

b.  Adjust the volume of the speakers using the “+” and “-“ buttons on the side of the right speaker.

4.  If using the microTargeting Controller and motor: without zeroing the motor, confirm that the microTargeting Controller 
 

motor rotates when the microTargeting Controller remote control is actuated.

B.  For each available channel, verify impedance check and stimulation functionality.

1.  Click on the Patient tab, apply the “Factory” profile, and then click on the Electrode tab.
2.  Click the “Impedance Check” button on the computer screen.
3.  Select a channel for testing using the on-screen dialog box. Confirm that the LED status indicator light for this channel is 
 yellow. 
4.  Plug the test electrode into the selected channel.
5.  Press the orange activate button on the Guideline 4000 LP+ stimulation remote control. 
6.  The measured impedance will be displayed on screen.  Confirm that the measured impedance of the test electrode is  
 

1±0.2 MOhm.

7.  Select “Done”.
8.  Click the ”Stimulation” button on the computer screen.  

Содержание 66-EL-LP

Страница 1: ...products MT LPP MT LPP APM 66 EL LP FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 www fh co com FHC Europe TERMOBIT PROD srl 42A Barbu Vacarescu Str 3rd Fl Bucharest 020281 Sector 2...

Страница 2: ...L011 57 Rev D1 2018 11 09 Page 2 of 11...

Страница 3: ...y Warnings and Cautions System Maintenance and Repair Handling Considerations Maintenance Repair System Disposal System Specifications Performance Accuracy Microelectrode Recording Stimulation Impedan...

Страница 4: ...complies with MDD Directive 93 42 EEC and legal responsibilities as a manufacturer are with FHC Inc 1201 Main Street Bowdoin ME 04287 USA Temperature limitation Storage conditions are between 0 C 40 C...

Страница 5: ...osal Return the Guideline 4000 LP to FHC for proper end of life disposal Please contact an FHC factory authorized representative to arrange a return System Specifications Performance Accuracy Microele...

Страница 6: ...patient isolated AC DC Coupled Bandwidth 0 5kHz Ranges factory configurable 0 to 16V 8 0V to 8 0V 0 to 8 0V 4 0V to 4 0V 0 to 1 6V 800mV to 800mV 0 to 800mV 400mV to 400mV 0 to 160mV or 80mV to 80mV...

Страница 7: ...speakers or Guideline 4000 LP and the patient simultaneously If using headphones use wireless headphones if contact with patient is possible WARNING The Guideline 4000 LP Multiple Portable Socket Out...

Страница 8: ...nfirm that a scrolling waveform is visible in the corresponding thumbnail window for each installed channel a If any channels have failed to initialize properly wait one minute for the LP to detect an...

Страница 9: ...nect the test electrode before connecting the LP to a patient 4 Load a user profile and enter patient information a Remove the test electrode if it is still attached to the LP b Click on the Patient t...

Страница 10: ...n using the orange activate button on the stimulation remote control If desired stimulation may be stopped by pressing the orange activate button again during fixed duration mode only WARNING Electrod...

Страница 11: ...e top center of the screen to view root mean square RMS voltage versus depth and mean firing rate MFR versus depth plots for all channels d Click on the Guidance tab at the top left of the computer sc...

Отзывы: