FHC 66-EL-LP Скачать руководство пользователя страница 4

L011-57 

   Rev. 

D1 

   2018-11-09 

   Page 

of 

11

Indications for use:

The microTargeting™ Guideline 4000 LP+™ neuromodulation targeting system is intended to assist in functional  
neurosurgical procedures where recording from and stimulation of brain motor and sensory neurons will aid in the  
placement of depth electrodes.

Intended use:

The Guideline 4000 LP+™ neuromodulation targeting system is intended to be used by a neurosurgeon, neurologist or  
clinical neurophysiologist to accurately position depth electrodes during functional neurosurgical procedures. 

Applicable Standards:

General Standard: EN60601-1 General Electrical Safety; Collateral Standards: 60601-1-1, 60601-1-2, 60601-1-4, 60601-1-4, 60601-1-6;
Particular Standard: EN60601-2-26; The Guideline 4000 LP+™ is a Class 1 medical device with a type BF applied part.

Symbol Key

WARNING / Caution, consult instructions 
for important cautionary information.

Consult the instructions for use.

In reference to “Rx only” symbol; this 
applies to USA audiences only

Caution

- Federal law (USA) restricts  

this device to sale by or on the order  
of a physician.

Medical device manufacturer, as defined 
in EU Directives 90/385/EEC, 93/42/EEc 
and 98/79/EC.

Telephone number

Indicates the catalog number so that the 
medical device can be identified.

Indicates the serial number so that a 
specific medical device can be identified.

Authorized Representative in the  
European Community

European Conformity. This device fully  
complies with MDD Directive 93/42/EEC  
and legal responsibilities as a manufacturer  
are with FHC, Inc., 1201 Main Street,  
Bowdoin, ME 04287 USA.

Temperature limitation: Storage conditions 
are between 0°C - 40°C (32°F - 104°F)

Humidity limitation: Not to exceed 95%

Atmospheric pressure limitation:  
between 500hPa and 1060hPa

Date when the medical device was 
manufactured.

Rx only

Rx only

0%

95%

j

n

Warnings and Cautions:

WARNING: The Guideline 4000 LP+ does no analysis, diagnosis, or monitoring for undesirable or dangerous physiological conditions.  

WARNING: The Guideline 4000 LP+ issues no alarms should undesirable or dangerous medical conditions arise.

WARNING: The Guideline 4000 LP+ must be operated by a person who has been trained by an FHC authorized representative and has 
read and understood the directions for use (DFU).

WARNING: The Guideline 4000 LP+ must be installed by an FHC authorized representative. 

WARNING: Do not use the Guideline 4000 LP+ system or any part of it as a diagnostic instrument.

WARNING: Potentially hazardous interactions between the Guideline 4000 LP+ and other patient connected equipment may occur.  
Review the documentation associated with all such equipment for relevant warnings and restrictions prior to use with the Guideline 
4000 LP+.

CAUTION:  Federal law (USA) restricts this device to sale by or on the order of a physician.

microTargeting™ and Guideline 4000 LP+™ are trademarks of FHC, Inc.

500hPa

1060hPa

Guideline 4000 LP+  
Breakout Box

Digital input

Analog input

Analog output

Guideline 4000 LP+  
Stimulation Remote Control

Audio jack

Type BF Applied Part

Microphone

Instructions for end of life 
disposal

0°C 
32°F

 +40°C 
 +104°F

Содержание 66-EL-LP

Страница 1: ...products MT LPP MT LPP APM 66 EL LP FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 www fh co com FHC Europe TERMOBIT PROD srl 42A Barbu Vacarescu Str 3rd Fl Bucharest 020281 Sector 2...

Страница 2: ...L011 57 Rev D1 2018 11 09 Page 2 of 11...

Страница 3: ...y Warnings and Cautions System Maintenance and Repair Handling Considerations Maintenance Repair System Disposal System Specifications Performance Accuracy Microelectrode Recording Stimulation Impedan...

Страница 4: ...complies with MDD Directive 93 42 EEC and legal responsibilities as a manufacturer are with FHC Inc 1201 Main Street Bowdoin ME 04287 USA Temperature limitation Storage conditions are between 0 C 40 C...

Страница 5: ...osal Return the Guideline 4000 LP to FHC for proper end of life disposal Please contact an FHC factory authorized representative to arrange a return System Specifications Performance Accuracy Microele...

Страница 6: ...patient isolated AC DC Coupled Bandwidth 0 5kHz Ranges factory configurable 0 to 16V 8 0V to 8 0V 0 to 8 0V 4 0V to 4 0V 0 to 1 6V 800mV to 800mV 0 to 800mV 400mV to 400mV 0 to 160mV or 80mV to 80mV...

Страница 7: ...speakers or Guideline 4000 LP and the patient simultaneously If using headphones use wireless headphones if contact with patient is possible WARNING The Guideline 4000 LP Multiple Portable Socket Out...

Страница 8: ...nfirm that a scrolling waveform is visible in the corresponding thumbnail window for each installed channel a If any channels have failed to initialize properly wait one minute for the LP to detect an...

Страница 9: ...nect the test electrode before connecting the LP to a patient 4 Load a user profile and enter patient information a Remove the test electrode if it is still attached to the LP b Click on the Patient t...

Страница 10: ...n using the orange activate button on the stimulation remote control If desired stimulation may be stopped by pressing the orange activate button again during fixed duration mode only WARNING Electrod...

Страница 11: ...e top center of the screen to view root mean square RMS voltage versus depth and mean firing rate MFR versus depth plots for all channels d Click on the Guidance tab at the top left of the computer sc...

Отзывы: