27
© Ferno Washington Italia ScoopExl - Rel.161115 - ENG
SCOOP EXL Stretcher
6 - MAINTENANCE
Maintenance
6.2 Disinfection
1. After each use, clean the surfaces of the stretcher with a
cloth and a non-abrasive surface disinfectant.
2. It is advisable to extend the leg section to the furthest
position and disinfect with the stretcher in an upright
position (foot end down).
3. Do not immerse the stretcher directly in disinfectant.
4. During disinfection, check that there are no signs of
wear and/or damage.
5. Dry with a cloth. Make sure that all the components of
the device are completely dry before reusing them.
It is advisable to wear suitable gloves (e.g. household or
disposable gloves) during disinfection.
6.3 Cleaning
1. Clean all surfaces of the stretcher with a cloth, warm
water and soap or a mild detergent soap.
2. If necessary, to remove stains, use a soft brush or a light
solvent, paying attention to labels.
3. Dry with a cloth. Make sure that all the components of
the device are completely dry before reusing them.
Do not wash with high pressure water
It is advisable to wear suitable gloves (e.g. household or
disposable gloves) during disinfection.
Important
The use of products containing bleaches, phenols and
iodine can damage the stretcher. Do not use products
containing these elements for cleaning and disinfecting
the stretcher.
6.1 Routine maintenance
The stretcher requires regular routine maintenance. It is
recommended to perform preventive maintenance on the
stretcher each year. When the device is put to particularly
heavy use, maintenance should preferably be carried out
on a more frequent basis by Ferno's technical personnel.
Repairs must be performed by Ferno's qualified personnel
using original spare parts.
If you discover signs of wear and damage, place the stretcher
out of use and immediately contact Ferno's Technical
Assistance Service (page 2).
Maintenance Periods
A
ft
er each use
W
hene
ver
nec
essar
y
M
on
thly
Disinfection
(page 27)
•
•
Cleaning
(page 27)
•
•
Inspection
(page 28)
•
•
•
Lubrication
(page 29)
•
Important
DO NOT WASH WITH HIGH PRESSURE WATER
Do not use high-pressure water to clean
the stretcher.
Important
After disinfection/cleaning, ensure the stretcher and its
components are fully and completely dry before use.