Fema M Series Скачать руководство пользователя страница 7

FEMA ELECTRÓNICA . SERIES M . M60-LC

7

Signal  

ranges

Accuracy

Max. input  

signal

0/100

mVdc

0.05% FS

30 V

0/30

mVdc

0.05% FS

0/20

mVdc

0.05% FS

0/15

mVdc

0.05% FS

0/10

mVdc

0.05% FS

0/5

mVdc

0.05% FS

Table 3 - Input signal ranges

1.9.2 

Signal ranges

The instrument works with 6 internal signal ranges and 

the active range is automatically selected when the in

-

strument is started. The selection depends on the value 

of the two parameters : ‘

Sensitivity

’ (

‘MV.V’

) and ‘

Excita-

tion voltage

’ (

‘V.EXc’

(see section 1.12.2)

Example

:

with a sensitivity configuration of 2.0000 mV/V 

and a configured excitation voltage of 10 Vdc, the instru-

ment selects the 20 mV input signal range, by calculat-

ing 2 mV/V x 10 Vdc = 20 mV.

The internal signal ranges available are shown below at 

Table 3

.

1.9.3 Number and type of cells accepted

The instrument accepts connection for up to 8 standard 

350 Ohms load cells. With a configured excitation voltage 

of 10 Vdc connect from 1 to 4 load cells. With a config

-

ured excitation voltage of 5 Vdc connect from 1 to 8 load 

cells.  For  load  cells  with  different  impedance,  calculate 

the current consumption for each cell, and the total must 

not exceed the maximum current the instrument can pro

-

vide.

In case of problems with the power or the signal provided 

by the load cells, the instrument provides three functions 

for  troubleshooting  purposes.  These  functions  allow  to 

access the signal input value (in mV), the excitation volt

-

age value at the ‘sense’ terminals (in Vdc) and the current 

provided to the cells (in mA). The operator can use this 

values to identify the cause of the problem. See section 

1.12.13

 for more information on how to access this val

-

ues in real time.

44

92

Panel 

cut-out

48

1.9.4 Mechanical dimensions (mm 

(in)

)

The instrument shows the value for the net weight, and can 

be configured to switch reading to gross weight and the actual 

value of the tare. The relation between them is :

• 

Net weight = gross weight - tare

Operator can access these values by configuring the fast ac

-

cess menu (key ‘UP (

5

)) 

(see section 1.12.13)

.

Reading the gross weight or tare values activate the front leds 

‘A’ and ‘B’ 

(see section 1.6)

.

1.9.5 

Gross weight, net and tare

1.9 Technical data (cont.)

96

(3.78 x 1.89 in)

(3.63 x 1.74 in)

16

8

75

(0.63

in)

(2.95

in)

(0.31

in)

Содержание M Series

Страница 1: ...nts tare and auto tare functions on power up functions stabil ity alarms configurable operator menu configurable led brightness Universal AC and DC power Up to 3 optional modules for output and contro...

Страница 2: ...lue in mV excitation current provided in mA and real excitation voltage see section 1 12 13 configurable Fast access menu key UP 5 with access to selected functions see section 1 12 13 Multiple displa...

Страница 3: ...configuration menu 19 1 2 Index 1 12 16 On power up configuration 19 1 12 17 Tools 20 1 12 18 Access to options configurations menu 20 1 13 Full configuration menu 22 1 14 Factory configuration 25 1 1...

Страница 4: ...ntrol options see section 2 the shipment also includes 1 pack of green signal terminals for each output and con trol option installed Figure 1 Set of units label 1 5 Additional information User s manu...

Страница 5: ...cells 2 3 4 cells or more con nect the sense wires to the electrical middle point of the power wires of all the cells see section 1 16 4 1 8 2 Connecting the cell to the ground Measuring with load ce...

Страница 6: ...G14 Configuration front keypad with 3 keys Front protection IP65 Outputandcontroloptions relay analog communications see section 2 Mechanical mounting panel connections plug in screw terminal housing...

Страница 7: ...ad cells For load cells with different impedance calculate the current consumption for each cell and the total must not exceed the maximum current the instrument can pro vide In case of problems with...

Страница 8: ...o the instrument and now the reading is 0 Kg The truck enters the loading area and when it leaves it is placed on the weighing system again Now the reading is 1 550 Kg This is the weight of the mate r...

Страница 9: ...ress the SQ key for 1 second to ac cess the configuration menu Inside the menu the SQ key acts as an ENTER It enters into the menu option selected and when entering a numerical value it validates the...

Страница 10: ...is particular load cell For the empirical configuration you will need access to two weights a low weight as small as possible it can be the cell without weight and a high weight as close as possible t...

Страница 11: ...d cell accepts any value between 0 and 999999 The decimal point will be shown in the posi tion configured at the Decimal point dP parameter De fault value is 1000 Example for a 5 Kg cell configure a v...

Страница 12: ...462 pounds Within the scale factor configure the multiplier to 220462 and the divider to 100000 The instrument is now configured to read in pounds The Stock units Stck function is provided to count la...

Страница 13: ...auto tare function The tare is automatically activated when reading is lower than the activation value and the fluctuation of the reading is lower than the counts defined at the stability band for th...

Страница 14: ...d the liquid is still moving inside the tank and this movement is reflected in weight and reading oscillations Alarm 2 is configured as stability alarm and activates when the liquid inside the tank is...

Страница 15: ...elay dEL 1 parameter config ure the delay to apply before the alarm is deactivated Delay starts to count once the setpoint is reached plus the hysteresis value Value from 0 0 to 99 9 seconds at the St...

Страница 16: ...er with higher values Increasing the strength of the filter slows the reading Value 0 disables the filter the Steps StEP function configures the reading to be done in steps of 1 2 5 10 20 or 50 counts...

Страница 17: ...bled the activation is per formed sequentially in the same order as the configuration menu first is the tare then the alarm unlock etc 1 12 11 Rear controls 1 12 12 Front key LE 3 The front key LE 3 c...

Страница 18: ...p function The On power up on Pu menu allows to define a series of functions to activate when the instrument restarts after a power loss Functions available are a delay on the activation of measure an...

Страница 19: ...he Exc voltage c EXc function allows to visualize the actual value of the excitation voltage measured between terminals sense and sense Value is offered in Vdc the Exc current c MA function allows to...

Страница 20: ...the Password func tion select on and enter the code The numerical code is asked when accessing the configu ration menu key SQ Functions configured to be ac cessible through the fast access menu are no...

Страница 21: ...FEMA ELECTR NICA SERIES M M60 LC 21 T h i s p a g e b l a n k...

Страница 22: ...l conf Load cell parameters Decimal point Sensitivity mV V Nominal weight Excitation voltage Field correction High weight correction Low weight correction Low weight value High weight value Stability...

Страница 23: ...Fixed digits Fix digits Memory of maximum Memory of minimum Rear controls Tare Alarmunlock Stock units Key LE Tare Alarmunlock Stock units Key UP fast access Gross weight Setpoint 1 Setpoint 2 Setpoi...

Страница 24: ...configuration Option 1 Access to the optional module installed at slot 1 Option 2 Access to the optional module installed at slot 2 Option 3 Access to the optional module installed at slot 3 1 14 Ful...

Страница 25: ...ff Locked alarms A LcK off Display dISP Fixed digits FIX d no fixed digits Average filter AVr 0 disabled Steps StEP 1 Left zeros LZEr off Memory of maximum MAX 199999 Memory of minimum MIn 999999 Rear...

Страница 26: ...powered from the instrument with power 10Vdc nominal weight 100Kg and 2mV V sensi bility Connect the load cell to the instrument see section 1 8 Configure the theoretical configuration of the load ce...

Страница 27: ...nt to measure a mil livolt generated signal at the laboratory there is a special configuration to apply Different from the load cell the millivolt generator is not a differential system and does not n...

Страница 28: ...he front filter at corner X and then insert clips A B C and D in this order Risk of electric shock Removing the front cover will grant access to the internal circuits Disconnect the input signal to pr...

Страница 29: ...tion required Instrument conforms to CE rules and regulations This instrument is warranted against all manufactur ing defects for a period of 24 months as requested by the European legislation This wa...

Страница 30: ...mitter B Collector C Not connected Figure 17 Connections for T1 transistor module The R1 module provides 1 relay output to install on Series M meters up to a maximum of 3 relay outputs in the same met...

Страница 31: ...t module 2 4 Module AO The AO module provides 1 isolated analog output configu rable for 4 20 mA or 0 10 Vdc to install on Series M me ters up to a maximum of 3 analog outputs in the same meter Output...

Страница 32: ...al_en pdf For additional information http fema es docs 4254_SERIES_M K S_OPTIONS_manual_en pdf Figure 22 Detail of the Modbus RTU communications module Power Opt 2 Opt 1 Opt 3 B A G B A G Signal B A G...

Страница 33: ...R4 or R6 module disables the alarm menus AL1 AL2 and AL3 of the instrument Choose special modules R2 R4 or R6 when a higher density of relays is needed Type of relay 3 contacts Com NO NC Max current...

Страница 34: ...ut front key pad To configure the instru ment remove the meter from the panel and remove the front filter Internal press buttons for configurationareaccessible Op tionally request the instrument preco...

Страница 35: ...1 THM benchtop housing Adapter for DIN rail mount for Series M K and C 4 2 Adapter DRA M Adapter 96x96mm for 96x48mm instruments 4 3 Adapter KA96 Wall mount housing To gether with the KIP protec tor...

Страница 36: ......

Отзывы: