Fema C Series Скачать руководство пользователя страница 11

FEMA ELECTRÓNICA . Series C . C40-D

11

 Measuring process signals

The instrument accepts the measure of process signals in 

4/20 mA and 0/10 Vdc. The instrument provides excita

-

tion voltage to power up transducers.

 Scaling

The instrument allows to scale the reading to 4 digits (9999 / -1999) 

with configurable  decimal point to any position. The ‘

second scaling

’ 

function can also be used 

(see section 1.25)

 

Maximum oversignal

‘Maximum oversignal’ is the maximum signal accepted by the instru

-

ment. Higher signal values may cause instrument damage. Lower val

-

ues are non destructive but may be out of accuracy specifications.

 

Response times

The response time to a signal step is 300 mSeconds, independent of 

the signal range selected.

 

Terminal 5 ‘multifunction’ - ‘Vexc’ or ‘External control’

To configure the +15 Vdc excitation voltage at terminal 5, set internal 

jumper ‘T’ at position 1-2 

(see section 1.9). 

Transducers with a con

-

sumption of up to 30 mA can be powered from this terminal.
To configure the ‘EK’ external contact function at terminal 5, set inter

-

nal jumper ‘T’ at position 4-5 

(see section 1.9). 

See section 

1.9 

for a list 

of available functions.

1.16 Process measures

 

Start-up, connections and jumpers

For instrument start-up follow the steps listed at section 

1.7

Signal 

connections are indicated at section 

1.5.

 Location for internal jumpers 

is indicated at section 

1.9.

See below a list of typical connections :

 

signal 4/20 mA active

 

signal 4/20 mA passive

 

signal 0/10 Vdc passive

 

signal (mA-)

+15 Vdc (mA+)

1 2 3 4 5

+Vdc

common

+15 Vdc

1 2 3 4 5

 

signal 0/10 Vdc active

 

mA +

mA -

1 2 3 4 5

+Vdc

common

1 2 3 4 5

Ranges 

 

of measure

Scale 

 

by default

Scalable

Jumper ‘S’ 

(see section 1.9)

Jumper ‘T’ 

(see section 1.9)

Accuracy 

(% FS)

Max.

 

oversignal

Connection

 

(

terminals)

Z

in

4/20 mA

passive 

(needs Vexc.)

0/100.0

de 9999  

a -1999

D

1-2

<0.15 %

25 mA

2 (signal)  

5 (Vexc)

4.7 Ohm

active

4-5

2 (mA+)  

4 (mA-)

0/10 Vdc

passive 

(needs Vexc.)

A

1-2 

<0.20 %

25 Vdc

2 (+Vdc)  

4 (comm.)  

5 (Vexc)

1 M

active

4-5

2 (+Vdc)  

4 (comm.)

Table 15 - Ranges of measure for process signals

1.17 Measuring frequency

 

How the instrument measures frequency

The instrument measures frequency from an AC voltage 

(Vac) or AC current (Aac) signal. The instrument detects 

each ‘0’ crossing of the signal, either ‘0 Vac’ or ‘0 Aac’. All 

available Vac and Aac signal ranges are accepted as frequency input 

signal.

 

How to configure the instrument to measure frequency

To measure frequency from a Vac signal, select the internal jumpers for 

the desired AC voltage range 

(see section 1.9)

, connect the signal for 

the selected voltage range 

(see section 1.10)

, and configure the instru

-

ment to measure frequency 

(see section 1.29.1)

The same applies to 

measure frequency from Aac signals. See example at section below.

 Scaling

The default resolution is 0.1 Hz. The instrument allows to scale the 

reading to 4 digits (9999 / -1999) with configurable  decimal point 

to any position. The ‘

second scaling

’ function can also be used 

(see 

section 1.25)

 

Maximum and minimum signal

Frequency signals below 15 Hz are measured as ‘0’. Frequency sig

-

nals  higher  than  100 Hz  are  out  of  accuracy.  Signals  higher  than 

1000

 Hz will read ‘display overflow’ ‘

d.oVr

’ error.

Ranges 

 

of measure

Scale 

 

by default

Scalable

AC signal

 

(see section 1.9)

Jumper ‘T’ 

(see section 1.9)

Response 

 

time

Accuracy 

(% reading)

15 Hz to 

100 Hz

0/100.0 from 9999  

to -1999

select Vac or 

Aac range

4-5

70 mSec. <0.15% of 

reading 

Table 16 - Ranges of measure for frequency

 

Example

To measure the 50 Hz frequency from a 230 Vac 

power line, select jumpers ‘GI’ for 600 Vac signal 

range 

(see section 1.9)

connect signal to termi

-

nals ‘1’ and ‘4’ 

(see section 1.10)

and configure 

‘frequency’  at  the  input  signal  configuration 

menu 

(see section 1.29.1).

Содержание C Series

Страница 1: ...ith 4 digits 9999 1999 Fast access function to alarm setpoints external control to activate predefined functions Eco mode for low consumption selectable dual scaling 5 levels of configurable brightnes...

Страница 2: ...23 Eco mode 13 1 24 External control 13 1 25 Second scaling 13 1 26 To open the instrument 14 1 27 How to operate the menus 15 1 28 Messages and errors 15 1 29 Configuration menu 16 1 29 1 Input signa...

Страница 3: ...terminals 1 and 4 Signals for 5 A current AC and DC must be con nected at terminals 3 and 4 All other signals must be connected between terminals 2 and 4 Terminal 5 is a multifunction terminal config...

Страница 4: ...r process signals 5 Vdc for potentiometers at terminal 5 see sections 1 5 and 1 16 Power supply power U 18 to 265 Vac dc isolation 1500 Veff isolation tested for 60 sec AC frequency 50 60 Hz consumpti...

Страница 5: ...Jumpers S Jumper T AC voltages and currents 600 Vac G I 4 5 200 Vac I 20 Vac A I 2 Vac B I 200 mVac C I 60 mVac E I 5 Aac I 20 mAac D I DC voltages and currents 600 Vdc G 4 5 200 Vdc 20 Vdc A 2 Vdc B...

Страница 6: ...steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vac ranges Veff Scale by default Scalable Jumper S see section 1...

Страница 7: ...ion 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vdc ranges Scale by default Scalable Jumper S see section 1 9 Jumper T see section 1...

Страница 8: ...ouple K E 4 5 100 1350 C 148 2462 F 2 tc 4 tc 3 Thermocouple J 100 1200 C 148 2192 F Thermocouple E 100 1000 C 148 1832 F Thermocouple N 100 1300 C 148 2372 F Thermocouple L 100 900 C 148 1652 F Therm...

Страница 9: ...0 see section 1 29 7 This pa rameter is associated to the specific model of probe installed Sensor Jumper S see section 1 9 Jumper T see section 1 9 Range in C in F Total error Connection terminals Cu...

Страница 10: ...n for internal jump ers is indicated at section 1 9 See below connections for NTC probe 1 14 Measuring with NTC probes NTC R25 configurable Jumper S see section 1 9 Jumper T see section 1 9 Range of m...

Страница 11: ...nnection terminals Zin 4 20 mA passive needs Vexc 0 100 0 de 9999 a 1999 D 1 2 0 15 25 mA 2 signal 5 Vexc 4 7Ohm active 4 5 2 mA 4 mA 0 10 Vdc passive needs Vexc A 1 2 0 20 25 Vdc 2 Vdc 4 comm 5 Vexc...

Страница 12: ...n 1 9 for a list of available functions Start up connections and jumpers For instrument start up follow the steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for int...

Страница 13: ...ree potential contact called Exter nal control To configure the second scaling 1 configure the internal jumper T to activate the EK external control function see section 1 9 2 associate the external c...

Страница 14: ...X and then insert clips A B C and D in this order The internal structure of the instrument is shown in the graphic be low Module Opt 2 connects to module Opt 1 Module Opt 1 connects to the display Opt...

Страница 15: ...t validates the num ber displayed Key UP 5 the UP 5 key gives access to the fast access menu Inside the menus it moves vertically through the different menu op tions At menus with numerical value entr...

Страница 16: ...able options Pt100 3 wire Pt100 2 wire Pt500 Pt1000 Ni100 Ni200 and Ni1000 For measures with 2 wires a manual compensation for the wire resistance error is available us ing parameter Offset reading oF...

Страница 17: ...ximum type alarm or minimum type activates when the dis play value is higher or lower than the setpoint value at the Setpoint SEt parameter enter the value for the alarm activation point This paramete...

Страница 18: ...vate a tare to the signal The actual signal will read 0 The calibration is not changed select parameter Decimal point 0 dP 0 to remove the decimal point on display select parameter Decimal point 1 dP...

Страница 19: ...compen sate for the possible error introduced by the resistance of the signal wires when measuring Pt probes Ni probes or resistances For more information see section 1 22 1 29 7 Menu Tools Option 1...

Страница 20: ...Ni1000 PTC and NTC For thermocouple probes resolution is fixed to 1 at the Degrees dEG menu select the temperature values to be read in C degrees Celsius or in F degrees Fahrenheit at the Alpha ALPh...

Страница 21: ...r 485 at Tools out 1 r 485 to have the serial output configuration at this menu entry Configure the address for the instrument parameter Address 1 to 247 Addr the bus speed parameter Speed kbps bAud...

Страница 22: ...c range 2 Vdc range 200 mVdc range 60 mVdc range 5 Adc range 20 mAdc range Process 4 20 mA range 0 10 Vdc range Thermocouples thermocouple K thermocouple J thermocouple E thermocouple N thermocouple L...

Страница 23: ...ss Memory of max Memory of min Setpoint 1 Setpoint 2 Degrees degrees Celsius degrees Fahrenheit Temperature resolution 1 resolution 0 1 resolution Second scaling Decimal point Display Low Display High...

Страница 24: ...elected at Tool Out 1 Analog output Display Low Display High Modbus RTU Address 1 to 247 Speed Kbps Format Password Version Factory reset Minimum Brightness Maximum Standard Alpha Alpha 385 Alpha 390...

Страница 25: ...ent Industrial CISPR 11 Instrument Class A Class B Group 1 For a detailed declaration see document www fema es docs 4152_CE Declaration_C40 D_en pdf Barber del Vall s January 2017 Xavier Junc Product...

Страница 26: ...portional to the reading of the instrument The mA output can be connected to work in active loop the module provides the power of the loop or passive mode the power of the loop is not provided by the...

Страница 27: ...420 see section 1 29 7 to inform the instrument that there is a M1 analog output module installed From this point you will need the calibration sheet provided together with the module M1 This sheet s...

Страница 28: ...the front filter Inter nal press buttons for configuration are accessible Optionally request the instrument preconfigured from factory Without front keypad 3 1 Option NBT O ring for IP65 front protec...

Страница 29: ...ll mount housing Together with the KIP protector offers a full IP65 protection For Series C of panel meters 4 4 WME wall housing Benchtop housing for Series C and of panel meters Handle with three sel...

Страница 30: ...FEMA ELECTR NICA Series C C40 D 30 Notes...

Страница 31: ...FEMA ELECTR NICA Series C C40 D 31 Notes...

Страница 32: ......

Отзывы: