F-900 Instruction Manual rev. 1/16/2014
1554 NE 3
rd
Ave, Camas, WA 98607, USA Phone (360) 833-8835 Fax (360) 833-1914
[email protected] http://www.felixinstruments.com
21
To avoid changing any configuration, press the STOP key to exit back to the Setup Menu.
Setup Sensor
The Setup Sensor Menu has three options,
Sensor Selection
,
Sensor Voltammetry
and
Closed
Loop Cleaning
. Sensor Selection allows the user to turn on/off different sensors of the F-900. Not
all units are built with all sensors, so see the Production Check Sheet at the end of the manual to see
which sensors are included in your unit. All units are equipped with a Standard, High Resolution
C
2
H
4
, ethylene sensor. Optional sensors available are Extended Range C
2
H
4
(ER), IRGA CO
2
and
Extended Range (ER) CO
2
(carbon dioxide), and O
2
(oxygen).
Sensor Selection:
Press the RIGHT arrow when Sensor Selection is highlighted to see the list of
sensors.
Use the UP/DOWN arrows to select the sensor to adjust and then the LEFT/RIGHT arrow
keys to switch between columns. Once in the column with the On/Off, use the UP/DOWN
arrow to turn the sensor On/Off.
If sensors are not included in your instrument, please set them to “Off”.
If not actively using the CO
2
sensor, set it to “Off” to conserve battery life.
Press SAVE to save any changes.
Press STOP to exit to the Setup Menu.
Sensor Voltammetry
should only be used if instructed by a Felix Instruments technician.
This is the electrode recovery mode and is a useful tool if the electrochemical cell has been
poisoned. However, running Sensor Voltammetry may cause the calibration to shift, changing the
data displayed on the screen.
Setup
Sensor
C
2
H
4
STD
On
C
2
H
4
ER
Off
CO
2
IRGA
Off
CO
2
ER
Off
O
2
Off