How GC Emulation Works
The patent pending Flow Injection Analysis method integrates the signal from the ethylene sensor
for 300 seconds. The resulting value, reported as area counts, is then compared against a user-
definable calibration point and an estimated concentration is determined. The user-definable
calibration point should be determined on a technician by technician basis as injection technique
will alter the result. Additionally, setting the calibration point near the expected ethylene
concentration will provide the most accurate estimation.
Improvements Compared to a Gas Chromatograph
1. No need for consumable bottled gases (N
2
, H
2
, air).
2. Portable.
3. No heating or ovens required.
4. Predicts ethylene concentration without chromatography.
GC Emulation Mode Operation:
1. Turn on GC Emulation mode in the Setup<Measure menu.
2. Enter the measure menu and wait for the instrument to stabilize.
Setup
Measure
Mode GC Emulation
SENSOR STABILITY
File
file01
Flow
200 mL
Initialization
Automatic
Environment Stabilizing
Please Wait...42%
Uptime 00:11 Bat 98%