![Felix Instruments F-900 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/felix-instruments/f-900/f-900_operation-manual_3916052017.webp)
F-900 Instruction Manual rev. 1/16/2014
1554 NE 3
rd
Ave, Camas, WA 98607, USA Phone (360) 833-8835 Fax (360) 833-1914
[email protected] http://www.felixinstruments.com
13
Monitor Mode
The large bold value on the top line is the digitally filtered value of the raw ethylene measurements.
The ethylene value may appear noisy, so using the filtered value helps stabilize the concentration
value. Next, the HUMIDITY, or relative humidity in percent, of the gas stream in the instrument is
displayed followed by the TEMPERATURE in degrees. The FLOW in mL per minutes is displayed.
The calibrated, recommended flow rate is 200 mL per minute. If the unit has a high resolution CO
2
sensor on board, the CO
2
value in ppm will be displayed below the ethylene value. If the unit has a
low resolution CO
2
sensor, or O
2
sensor on board, the value will be displayed in percentage (%). If
the unit is equipped with a low resolution ethylene sensor, the value will continue to be displayed
in large, bold font, but the range will be extended up to 100 ppm.
If Autosave is enabled, “saved” will flash at the top of the display each time the data is
automatically saved to the file.
Press SAVE to save the currently displayed data to the SDcard. “Saved” will briefly flash on
the top of the display.
Press the LEFT arrow or STOP key to stop the measurement and exit to the Sensor Status
display.
Press the DOWN arrow to enter GRAPH Mode.
To use the F-900 in Monitor Mode to continuously monitor fruit storage rooms or other locations,
or to run the instrument for days or weeks at a time, connect the mini-USB cable to the charger
cable and to the connection port on the front panel. Plug the charger into an electricity source.
MONITOR
3.26 ppm
Humidity 47.0%
Temperature 28.7C
Flow 200 mL
Uptime 01:17 Bat 95%