FEASA LED ANALYSER LOW LIGHT
FUNCTIONAL VERSION
Daisy Chain Mode
10.
To connect to Analyser 2 (s/n F461) send:-
busfree
OK
Response from the Analyser
busgetF461
OK
Response from the Analyser
The host computer is now connected to Analyser 2 (s/n F461).
This can be verified by requesting the Serial Number:-
getserial
F461
Response from the Analyser
All the LED Analyser commands can now be directed to Analyser 2.
11.
To connect to Analyser 3 (s/n F201) send the following:-
busfree
OK
Response from the Analyser
busgetF201
OK
Response from the Analyser
12.
It is possible to instruct all Analysers in the chain to capture LED data simultaneously.
This makes programming easier and saves time. There are six capture commands in
total but only one can be used at a time. The commands are
busc, busc1, busc2,
busc3, busc4 and busc5
.
Busc
uses the automatic capture range whereas
busc1 –
busc5
use pre-selected manual ranges.
13.
If you are have difficulty reading from Analysers at the end of your chain please verify
your voltage to each of these units is 5V.
67