FEASA LED ANALYSER LOW LIGHT
FUNCTIONAL VERSION
General Commands
EOT -
End of Transmission
Character
Transmit
Receive
enableeot
disableeot
See details below
Description
Every command sent to the LED Analyser or response received from the Led Analyser, ends
with a set of characters that indicates that the command or response has finished. Those
characters are CR + LF, which are ASCII characters 13 and 10 respectively. Alternatively,
responses from the LED Analyser can also have an end-of-transmission format like CR + LF +
EOT, which are the ASCII characters 13, 10 and 04 respectively. CR + LF are always present,
but character EOT can be enabled or disabled using the commands ENABLEEOT and
DISABLEEOT. Having EOT character enabled can be extremely useful in multi-line commands,
since this allows to differentiate between the end of a line and the end of a transmission.
The EOT command is ENABLED by default. To disable EOT simply send the command
disableeot using Feasa Terminal software. The Analyser will maintain the EOT state even when
powered off. To enable the state just send the command enableeot.
Information Received back from the Analyser using the
getstatus
command is as follows:
Hardware: Feasa xx-y
Serial number : Xyyy
Firmware Version : Iyyy
Exposure factor: 05
Number of averages: 01
Capture : Manuel 04
Number of Fibers : 05
End of Transmission Character:
Enabled
Selected calibration set: 0
user calibration date: dd/mm/yyyy
Factory calibration date: dd/mm/yyyy