FEASA LED ANALYSER LOW LIGHT
FUNCTIONAL VERSION
User Calibration Mode
setCALIBRATION## - Select the Calibration Set to use
Transmit
Receive
setcalibration#
putcalibration#
OK
OK
Where:
#
represents the set value (# can be 0 to 9).
Description
This command will select which calibration set you are accessing. The Led analyser is capable
of storing fifty one (51) separate user Calibration sets on the board.
The command
SETCALIBRATION#
stores the Calibration sets both on RAM and on EEPROM
(stays after power cycle but limits the number of
set
command operations to ~ 100K).
The command
PUTCALIBRATION#
only stores them on RAM and the analyser will default back
to what values were stored in the EEPROM after a power cycle.
The default calibration set is 0
Example:
If you want to select set No.3
setcalibration3
OK
Example:
If you want to select set No.3
putcalibration3
OK
The command
GETCALIBRATIONSET
will return the number of which stored Calibration set of
values are currently selected.
Example:
calibrationset3
52