4.3
(a) Mode selection
(b) Fixed Trigger To Laser value
Figure 4.18: “No Delay” mode
4.3.5
High repetition rate
In normal use, there is a different acoustic wave for each diffracted optical pulse. This generally
prevents using the Dazzler
TM
with repetition rates higher than 1
/T
Xtal
(time needed to refresh
the acoustic pulse in the crystal). For HR25 Dazzler
TM
systems, this limit corresponds to 30
kHz repetition rate.
It is however possible to run a Dazzler
TM
at higher optical pulse repetition rates, at the expense
of a lower pulse shaping capability and some small phase variations between consecutive pulses.
This mode is called “High Repetition Rate”, and is switched on automatically when:
1
RepRate
+
Trigger to Laser
≤
(
T
p
+
T
Xtal
+
d
M in
)
(4.4)
When “High Repetition Rate” mode is active, a blinking indicator appears in the upper right
corner of the
Trig&Mode
panel. The number of optical pulses diffracted for each acoustic cycle
is then shown in the
N
indicator. The graph along with its legend are updated (
Figure 4.19: ”High Repetition Rate” mode - (HR 25 Dazzler
TM
model typical values)
The acoustic cycle duration in this mode is given by :
τ
Ac
=
N
·
1
/
(
Repetition Frequency
)
(4.5)
V3.00 - 8