DAZZLER
TM
system manual Part I : installation & operation
3.3
Figure 3.4: Details of the Dazzler crystal
3.3.1
Frequency Calibration
The Dazzler has qto be frequency calibrated to achieve accurate pulse shaping. Use an ampli-
tude spectrum with a narrow hole (example on
Figure 3.5: Beam position on the input face
Figure 3.6: Example of spectrum
with hole
Feed the diffracted beam into a well calibrated spectrometer. Verify the wavelength at which
the hole feature appears. If a difference
dλ
occurs with the programmed value and this difference
is significant (i.e. larger than 10nm), there is a setup problem, a non exhaustive list of possible
causes is:
•
the input light polarisation orientation is incorrect
•
the autocollimation is incorrect (e.g one degree error)
V3.00 - 8