ASSEMBLY OF THE CABINET ON THE TABLE (optional) 

 

Содержание FlowFAST H

Страница 1: ...20EN Rev 09 12 2012 ORIGINAL INSTRUCTIONS Commercial office Via Merendi 22 20010 Cornaredo MI Tel 39 02 93 991 92 Fax 39 02 93 991 608 E mail info faster dgroup it OPERATING MAINTENANCE MANUAL FlowFAS...

Страница 2: ...NT OF HEPA FILTER BY TECHNICAL ASSISTANCE PERSONNEL 22 7 D REPLACEMENT OF MOTOR FAN BY TECHNICAL ASSISTANCE PERSONNEL 22 7 E REPLACEMENT OF FLUORESCENT LAMPS BY USER 23 7 F REPLACEMENT OF U V GERMICID...

Страница 3: ...g Microbiology Preparation of culture media Preparation of ophtalmic solutions TPN Preparation Assembly of equipment Floriculture Furthermore FlowFAST H cabinets fulfil the armonized standards EN 6101...

Страница 4: ...all the cabinet in a well ventilated room with a low degree of dust The distance between the cabinet and the room ceiling should be at least 10 cm Place the cabinet away from doors and windows which m...

Страница 5: ...tipulated by local legislation insert upstream of the power cable an automatic protection overload switch provided with a differential relay with a rated switching voltage no greater than 30 mA The ri...

Страница 6: ...0 1495x734x600 1800x734x600 Weight Kg 95 128 175 188 Noise level dB A 58 58 58 58 Lighting level Lux 800 800 800 800 Main voltage V 230V AC 2P T 230V AC 2P T 230V AC 2P T 230V AC 2P T Frequency Hz 50...

Страница 7: ...motor fan placed in the upper part of the cabinet through a pre filter which eliminates the dust in order to protect the motor fan and preserve the HEPA filter from premature clogging The air is subs...

Страница 8: ...ipulated in the working chamber The cabinets meet ISO 14644 1 Class 5 requirements 4 B CONTROL and REGULATION SYSTEMS The FLOWFAST H cabinet is provided with a manual electronic regulation system or i...

Страница 9: ...e PCB for the automatic regulation is installed OPTIONAL insert the password using the small numbers in the right lower corner of the keys and press SET Default password is 5 4 When the password is ty...

Страница 10: ...when enabled the display shows U V on The U V lamp switches on only if the cabinet is off the lighting of the cabinet is off and the front window is completely closed By pushing the relevant key you...

Страница 11: ...lay shows the notice es MOT LAF XX max 100 Operating Time Shows the operating time of the cabinet from the moment when the main switch is positioned on I The LCD will display for example WORK TIME XXX...

Страница 12: ...oing back to the upper level 12 LINE The green mains light switches on if the unit is connected to the mains and the line is live 13 DISPLAY optional Rearlight liquid crystal LCD display composed of 2...

Страница 13: ...p the clock OPERATOR MENU Change Password OPERATOR MENU Alarms display OPERATOR MENU UV residual time STAND BY MODE PASSWORD ESC UP arrow OPERATOR MENU Timer set up OPERATOR MENU Alarm clock set up OP...

Страница 14: ...IDUAL LIFETIME and press SET 8 key the display will show FILTER 1 RESIDUAL LIFETIME set XXXX where XXXX shows the number of the hours set for lifetime of the FILTER 1 Use the up and down arrow keys to...

Страница 15: ...hoice the desired menu select PASSWORD CHANGE and press SET 8 key the display will show CURRENT PSW digit the present Password then press SET key SET UP PASSWORD PSW digit the new Password then press...

Страница 16: ...the buzzer start to ring and the display show this message ALARM CLOCK ALARM Press the ESC key to silence the acoustic signal TIMER SETUP use UP DOWN arrow keys 5 6 to choice the desired menu select T...

Страница 17: ...n XXXX Use the left and right arrow keys to select the desired parameter and set it using the up and down arrow keys Then press the SET key to confirm the data The display shows UV TIMER DATE XXXX Use...

Страница 18: ...ry Appropriate disposal of this product will help to prevent potential negative effects on health and environment and to promotes re use and or recycling of materials of the equipment The improper dis...

Страница 19: ...work DO NOT eat drink or smoke in the work zone AVOID substances that release explosive vapours In addition when working with the cabinet AVOID the introduction of extraneous material quick movements...

Страница 20: ...n the lighting by pushing the key The cabinet is ready to work Wait 5 minutes before starting any operation see chapter 5 During this period of time introduce ALL and ONLY the material indispensable f...

Страница 21: ...ct for stainless steel sheets Clean sterilize the internal work surface with an efficacious detergent and a bactericide e g ethanol phenolic compositions aldehydes quaternary ammonium salts etc IMPORT...

Страница 22: ...use of tools between the HEPA filter and the frame 4 For re assembly reverse the above instructions taking care a Don t tighten nuts 13 too much b To centre the filter in its housing N B The new gask...

Страница 23: ...ter 5 For reassembly reverse the above instructions 6 Re connect the cabinet 7 F REPLACEMENT OF U V GERMICIDAL LAMP optional by user See chapter 13 diagram for maintenance operations 1 Switch the cabi...

Страница 24: ...PCB PW LAF 1 1 1 1 V30000005200 LAF touch sensitive keyboard 1 1 1 1 V20000006020 18 W 84 fluorescent lamp 2 V20000006030 30 W 84 fluorescent lamp 2 V20000006040 36 W 84 fluorescent lamp 2 2 V20000006...

Страница 25: ...ontamination of the product but protection of the environment The main motor fan does not work Check the terminal voltage of the power card of the main motor fan Check F7 fuse on the power card Replac...

Страница 26: ...by manual regulation of the air speed or automatic if in the version with automatic regulator and digital anemometer OPTIONAL automatic regulation with digital anemometer to keep the velocity of air...

Страница 27: ...table 2 Place the work surface 2 on the main unit 1 in the lower part checking that the holes of the main unit and of the work surface fit together Fasten them with the relevant screws and washers 3...

Страница 28: ...tle the cabinet in three sections 1 main unit 2 work surface 4 safety glass side panels 21 22 23 shutter slide guide and stirrup for shutter hooking For disassembling operations please refer to chapte...

Страница 29: ...x 85 x 145 5 cm 1 cm thick FLOWFAST H 15 202 5 x 82 x 147 cm 1 cm thick FLOWFAST H 18 202 5 x 82 x 147 cm 1 cm thick Steel strap and clips During transport pay attention to maintain the package in ver...

Страница 30: ...in the instruction manual of which the most important are listed again below installation in a place which does not conform to the manufacturer s recommendations wrong power voltage poor earthing use...

Страница 31: ...2 Work surface 3 Screws 4 Glass 5 Screws 6 Fastener frame 7 Junction 8 Motorblower screws 9 Motorblower 10 Pre filter 11 Screws 12 Rear panel 13 Nuts 14 HEPA filter 15 Lamp holder 16 Nuts 17 Lamps 18...

Страница 32: ...a D GROUP company 00020EN Rev 09 12 2012 32...

Страница 33: ...a D GROUP company 00020EN Rev 09 12 2012 33...

Страница 34: ...a D GROUP company 00020EN Rev 09 12 2012 34 14 FLOWFAST H FRONT and SIDE DIAGRAMS 14 A FRONT DIAGRAM A B FlowFAST H 09 885 945 FlowFAST H 12 1190 1250 FlowFAST H 15 1495 1555 FlowFAST H 18 1800 1860...

Страница 35: ...a D GROUP company 00020EN Rev 09 12 2012 35 14 B SIDE DIAGRAM...

Страница 36: ...a D GROUP company 00020EN Rev 09 12 2012 36 15 CABINET MOUNTING...

Страница 37: ...a D GROUP company 00020EN Rev 09 12 2012 37...

Страница 38: ...a D GROUP company 00020EN Rev 09 12 2012 38 16 ASSEMBLY of SUPPORT TABLE...

Страница 39: ...a D GROUP company 00020EN Rev 09 12 2012 39 17 ASSEMBLY OF THE CABINET ON THE TABLE optional...

Страница 40: ...a D GROUP company 00020EN Rev 09 12 2012 40 18 SIDE DIAGRAM WITH U V LAMP optional...

Страница 41: ...company 00020EN Rev 09 12 2012 41 19 SENSORS LIST PLUG Software name Description J4 CPULAF S0 Air flow sensor J3 CPULAF S1 J2 CPULAF S2 J3 PWLAF S3 Magnetic sensor for U V J2 PWLAF S4 Proximity switc...

Страница 42: ...a D GROUP company 00020EN Rev 09 12 2012 42 20 WIRING DIAGRAM...

Страница 43: ...a D GROUP company 00020EN Rev 09 12 2012 43...

Страница 44: ...a D GROUP company 00020EN Rev 09 12 2012 44...

Страница 45: ...a D GROUP company 00020EN Rev 09 12 2012 45...

Страница 46: ...the harmonisation of the laws of Member States relating to electrical equipment designed for use within certain voltage limits and with the following standards ISO 14644 1 Cleanrooms and associated co...

Отзывы: