Software Description
; bit 10: GO Low at Stop
; bit 11: Clear before triggered start
digval=0
; 0..255 DIG I/O Output value
trigger=6
; bit 1..4 control_trigger for scope mode.
; bit 0: TRG_EDGE_DOWN
; bit 1: TRG_EDGE_UP
; bit 2..4: TRG_SRC,0=none (softw.), 1..4=CH1..4, 5=extern
; bit 5: 0=normal, 1=auto
; bit 6: single shot
mcsmode=0
; bit 16: mcs mode
; bit 17: 2mcs mode
; bit 18: sequential mode
; bit 19: ext. clock
cycles=1
; cycles for sequential mode
mcspol=10
; bit 0: start with rising edge
; bit 1: stop1 rising edge
; bit 2: stop2 rising edge
; bit 3: ch adv rising edge
; bit 4: FAST NIM
; bit 5: 4k7 MCS Inputs
svapol=f
; for external trigger and SVA:
; bit 0: trigger / ch1 with rising edge
; bit 1: ch2 rising edge
; bit 2: ch3 rising edge
; bit 3: ch4 rising edge
; bit 4: FAST NIM
; bit 5: 1=4k7, 0=50 Ohm
dwellunit=0
; choice of units in dialog 0=ns, 1=usec, 2=msec, 3=sec
dwelltime=10
; in units of 10 ns
swpreset=10
; sweep preset value
timepreset=10.000 ; time preset
unipolar=1
; unipolar or bipolar (0)
event_cnt_mode=1
; count only stored events
fifosize=0
; 0=128 MB, 1=32 MB, 2=8 MB
sca1=ffff
; LOWORD: upperlevel SCA 1, HIWORD: lowerlevel SCA 1
sca2=ffff
sca3=ffff
sca4=ffff
phafit=0
; bit0..1 chn1: 0=max, 1= fit, 2 flattop, 3 reserved
; bit2..3 chn2
; bit4..5 chn3
; bit6..7 chn4
fitwidth1=29
; LOWORD: fitwidth gaussian, HIWORD, pulsewidth flattop
fitwidth2=29
fitwidth3=29
fitwidth4=29
memsegs=aa
; Bit0..1: SEGMENTS_CH1, 0..3 = 1x32k/2x16k/4x8k/8x4k segments
; Bit2..3: SEGMENTS_CH2
; Bit4..5: SEGMENTS_CH3
; Bit6..7: SEGMENTS_CH4
ctm8=125
; coincidence time in units of 8 ns
mwidth1=0
; flattop: range for Mean calc LOWORD upper, HIWORD lower
mwidth2=0
; LOWORD: Bit0..13 meanrange, Bit(15,14) Method 0=left, 1=cft, 2=right
mwidth3=0
; HIWORD: roileft
mwidth4=0
gate=0
; bit 0..3: GATE_ADC1..4, 1=Gate must be active for pulse processing
; bit 4..7: GATE_POL1..4, 1=active high
; bit 8..11: REJECT_ADC1..4, 1=use GATE as reject, e.g. pile-up detected,
; cor. GATE_ADCx must be 0
; bit 12..15: SVA mode
ctlsca=0
; bit 0..3: SCA_CH1, 0=SCA, 1=Over-Threshold, 2=GATE Input, 3=MCS Input
; bit 4..7: SCA_CH2,
F
ComTec GmbH
63
Содержание MCA4A
Страница 2: ...2 F ComTec GmbH...
Страница 8: ...8 F ComTec GmbH...
Страница 71: ...Software Description F ComTec GmbH 71 Fig 4 35 Control Panel of the demo VI for LabVIEW...
Страница 96: ...MPANT Software 96 F ComTec GmbH Fig 5 29 MPANT with four systems enabled...
Страница 106: ...Appendix 106 F ComTec GmbH...
Страница 107: ...Appendix 7 8 Personal Notes F ComTec GmbH 107...