Model ULC Series
06012021ULC,Rev i
Page 14 of 54
Alarm Silence
This icon will be displayed during an active alarm. Touching this Alarm Silence
Icon will silence the audible alarm for a period of time set up by the user. Should
the alarm condition remain active, the audible will sound again at the end of the
delay period.
Warning Icon
This icon will be displayed during an active warning. The “Warning Icon” is visible
when a warning message is active. Touching this “Warning” icon will display the
active warning message(s). A warning informs users that an Alarm Condition
may occur if the warning conditions is not resolved within the programmed delay
period. Example: door open warning with system in a defrost mode.
Alarm Icon
This icon will be displayed during an active alarm. The “Alarm Icon” is visible
when an alarm condition is active. Touching this “Alarm” Icon will display the
active alarm condition(s). Action is required by the end user to resolve the Alarm
condition. Example: door open alarm, system failure alarm.
Service
Icon
This icon will be displayed to indicate the system is due for maintenance or
calibration.
Note: Operator privileges or higher will be required to make changes
to any of the controller settings. The access will return to view only if
the control is left unattended for more than 5 minutes.