background image

EN

Error Codes

16

Wile  moisture meter is equipped with error messages to guide the user to get the 

most reliable results, as well as to ensure that the meter is in good condition.

Error code

Cause

Solution

Critical Error 

The meter must be submitted to an autho-

rized service center

7

Weight not detected

Repeat test / Do weight test

8

Taring result is too high

Do weight test and check the scale

9

Weight measurement uncertain / 

meter shook during the measure-

ment

Keep the meter stable, 

measurement uncertainty has increased 

10

Moisture measurement result stabi-

lize slowly

Happens to very wet samples

11

Weight measurement too uncertain 

/ meter shook during the measure-

ment

Keep the meter stable, measurement uncer-

tainty has increased

12

Temperature of the sample lower 

than 2 C / Measurement uncertainty 

is high

The tester is most accurate when Rice and 

tester temperature is between 16C and 32C

13

Temperature of the sample above 

50 C / Measurement uncertainty is 

high

The tester is most accurate when Rice and 

tester temperature is between 16C and 32C

14

Large temp. diff. between the Rice 

and test cell / Measurement uncer-

tainty has increased

For better accuracy, Rice and meter test cell 

should be at similar temperatures

15

Not on a suffciently level surface, 

meter warns with sound alarm

Keep meter stable during measurement

16

Weight result is over 330 g

Check that leveling handle did not stay 

above the test cell/ Do weight test

17

Weigh result is under 50 g

Check that leveling handle did not stay 

above the test cell/ Do weight test

18

Hectoliter weight is 20% higher than 

average 

Measurement uncertainty has increased, Do 

weight test

19

Hectoliter weight is 20% lower than 

average  

Measurement uncertainty has increased,      

Do weight test 

20

Test cell is not cleared before the 

measurement

Clear test cell

Содержание Wile Rice

Страница 1: ...Moisture meter EN Operators manual ...

Страница 2: ...nsation guarantees accurate results The meter is equipped with a patented integrated Rice leveling system Excess Rice goes into the compartment around test cell keeping working area neat during whole measuring process Read this manual carefully to learn how to operate this unit correctly 1 METER 2 CARRYING CASE 3 USB CABLE 4 BATTERIES 5 MANUAL Box Contents 1 4 5 3 2 10 Moisture meter EN Operators ...

Страница 3: ... batteries 4 0 Moisture is below the lower limit for this grain 28 0 Moisture is above the upper limit for this grain Error code MEASURING PROCESS Wile guides user step by step directly on the screen during the whole measur ing process 1 Place hopper on tester Press and release ON OFF button to turn on tester After power up the display shows the results of the last Rice tested 2 Select Rice using ...

Страница 4: ...rds the back of the unit and back removing excess grain 6 When the handle is back to starting position the meter will automatically start measuring Results will appear on the screen within seconds FUNCTIONS AND SETTINGS Settings and functions can be browsed and modified at the settings menu RICE OFFSET GRAIN OFFSET Each Ricecan be individually adjusted by 10 0 in increments of 0 1 to match the res...

Страница 5: ...ed AUTO OFF The meter will turn off automatically after a selected time 30s 1min 5min 10min 20min AVERAGE CALCULATION The meter automatically calculates the average of the previous measurements In settings menu you can adjust the number of results 3 6 or 9 measurements for the average calculation The average can also be reset BACKLIGHT ON OFF To extend battery life you can turn display backlight o...

Страница 6: ...ate it on the startup screen automatically but measuring is still possible Low battery voltage does not affect the measurement results WEIGHT TEST Scale function can be checked at weight test mode Use known weight sample up to 500 grams of weight Follow instruction on the screen of your meter INFO Info section shows software version details ...

Страница 7: ...test cell MAINTENANCE Use a soft clean cloth to clean device Do not use strong detergents and do not allow liquids inside the meter Keep the meter in a dry place preferably at room temperature If the meter is not used for a long time remove the batteries BATTERY REPLACEMENT The meter uses four AA LR6 batteries Replace all batteries at the same time Temperature sensor ...

Страница 8: ... is high The tester is most accurate when Rice and tester temperature is between 16C and 32C 13 Temperature of the sample above 50 C Measurement uncertainty is high The tester is most accurate when Rice and tester temperature is between 16C and 32C 14 Large temp diff between the Rice and test cell Measurement uncer tainty has increased For better accuracy Rice and meter test cell should be at simi...

Страница 9: ...ner will repair or replace the defected product and return it as soon as possible The warranty does not cover any damages that are caused by incorrect or careless use of the product installation that does not correspond to the provided instructions and other damages which may arise due to causes beyond the control of the manufacturer The liability of the manufacturer is limited to the pRice of the...

Страница 10: ......

Страница 11: ......

Страница 12: ...Jusslansuora 8 04360 TUUSULA FINLAND Tel 358 9 7744 970 info farmcomp fi www wile fi Copyright Farmcomp Oy 2019 All Rights Reserved 99209110 ...

Отзывы: