DNA
TM
System Instruction Manual
D00981262
Revision A, May 2015
22
Tolerance
Each test consists of profiles that contain steps. When the tolerance is set to 5, the
software will go to the next step when the temperature is within 5 degrees of the set
point temperature. For example, if the set point is 120°F, the next step will begin
when the temperature increases to 115°F.
Over Temp
Over Temp is a global setting that applies to all machines configured in the DNA
TM
system. Each machine has a safe temperature setting, which the user cannot set
temperature above. Ensure that the machine is set to a safe temperature setting
when running a test to avoid the Over Temp from stopping the test from reaching
the maximum temperature setting.
•
DNA
TM
machines have safety settings of 300°F and will not run tests above
this temperature.
•
Fann Thermocups have safety setting of 190°F to prevent overheating.
Settle Time
This time (in minutes) defines how long the temperature remains stable within the
tolerance limit before the system proceeds to the next step.
4.7
Updating the Software
Click the
Update
button on the
Main Menu
.
Follow the instructions shown on the screens until the update is complete.