DNA
TM
System Instruction Manual
D00981262
Revision A, May 2015
18
•
To enter
SETUP
mode, press and hold STOP, and then turn on the power.
•
Wait for the
SETUP
menu to appear.
•
Press the blue key 4 to enter the address in the number range 001 to 255.
RheoVADR
®
units come with the address set to 110 in manufacturing. Each
RheoVADR
®
connected to the DNA
TM
will need a unique address.
Model 741 Communication Address
Eurotherm Display
•
Hold the
Page
button (~ six seconds) until
Level 3
appears.
•
Quickly press the
Raise
(up arrow) button until
Config
appears, and then wait
until
Code
appears.
•
Use the up or down arrows to set the
Code
to 7. Wait until
Config
appears.
•
Press the
Page
button multiple times until
Comms
appears.
•
Press the
Scroll
button until
id
appears. Then set the id to
rs485
using the
arrow keys.
•
Press the
Scroll
button until
ADDR
appears. Then set the desired address using
the arrow keys.
•
Press the
Scroll
button until
BAUD
appears. Then set the baud rate to 19.20
using the arrow keys.
•
Press and hold the
Page
button until
Config GOTO
appears. Then press the
arrow keys to go to
Level 1
.
•
Wait for the Eurotherm display to return to the main screen (shown above).