Instruction Manual for the Delivery Table FREYA 3 

Page 4 

press  your  hand  against  the  table  elements  near  the  moving  elements  of  the 
structure. 

  If the patient lies on the table without being supervised by the medical staff, then 

the table should remain in the lowest position. 

  If  the  patient  is  not  attended  to  by  the  medical  staff,  lock  the  table  functions 

activation of which may pose a threat to the patient safety. 

  If  there  is  a  risk  that  the  patient's  condition  (e.g.  disorientation  due  to  medical 

reasons  or  her  clinical  state)  might  lead  to  trapping  the  patient,  then  the  lying 
surface  ought  to  remain  flat,  in  the  absence  of  supervision  (unless  otherwise 
advised by the medical personnel, or under special/particular circumstances). 

  Avoid safety risks which may result from the improper use of power cables/cords, 

e.g.  tensioning  the  cables/cords,  their  squeezing  in  between  the  moving  parts, 
running over them. 

  Do not roll the delivery table across electrical cables! 

  The  patient's  weight  and  additional  equipment's,  mounted  at  the  table,  may  not 

exceed 40 st (250 kg). 

  Maximum permissible load of the extended lying surface: up to 27 st (170 kg). 

  Obligatorily exercise caution while adjusting the extended lying surface in relation 

to  the  permanent  one,  as  the  distance  between  the  permanent  frame  and  the 
extended one is too small (if the pressure knob is used). In extreme cases, it may 
lead to crushing the patient's or user's fingers.  

  Maximum permissible working load of the foot support: up to 500 N. 

  If the device is to be used as a standard bed, mount the footboard. 

  Maximum  permissible  working  load  of  the  calf  support,  mounted  in  the  fastening 

holder, on the fastening strip: up to 5 st (30 kg). 

  Maximum  permissible  working  load  of  the  hand  holder,  mounted  in  the  multi-

position holder, on the fastening strip: up to 500 N. 

  Maximum  permissible  working  load  of  all  the  infusion  rod  hooks:  up  to  0.5 st 

(3 kg). 

  The  delivery  table  is  intended  for  patients  whose  height  does  not  exceed  79 in 

(200 cm).  

 

Start-up, operation and use notes 

  During normal use, the delivery table must obligatorily be connected to the mains.  

  The bed, during use, should not be in the place which obstructs its disconnection 

from the mains! 

  If  you  use  the  table  near  high-frequency  medical  apparatus  and  defibrillators, 

carefully  observe  recommendations  contained  in  instruction  manuals  of  this 
equipment. Improper use may result in dangerous accidents. If the patient touches 
the metal table parts or equipment, she runs the risk of serious burning. 

Содержание FREYA 3

Страница 1: ...INSTRUCTION MANUAL Delivery table type LM 03 FREYA 3 Serial no Revision 2 00 Publication date 26 03 2015 FAMED YWIEC Sp z o o Annexes 2 3 4 Index of Spare Parts...

Страница 2: ...e essential requirements contained in Annex I to the directive and Act on medical devices The compliance procedure has been performed acc to Annex VII to the directive Manufacturer FAMED YWIEC Sp z o...

Страница 3: ...ranty null and void Before any repairs disconnect the device from the mains Safety notes Below is the warning Warning Follow the instructions for safe use This warning can be found on parts and mechan...

Страница 4: ...ly exercise caution while adjusting the extended lying surface in relation to the permanent one as the distance between the permanent frame and the extended one is too small if the pressure knob is us...

Страница 5: ...r underneath must be free of obstacles During the table movement avoid collisions Exercise caution while removing and mounting the permanent lying surface moulding Application of excessive force may l...

Страница 6: ...8 ARMCHAIR POSITION OF THE DELIVERY TABLE CARDIAC CHAIR 18 3 9 QUICK LEVELLING OF THE THIGH AND BACKREST SECTIONS CPR FUNCTION 18 3 10 POSITION 0 18 3 11 EXTENDED LYING SURFACE FRAME POSITION ADJUSTME...

Страница 7: ...used The user is obliged to ensure that the device is operated in compliance with its intended purpose and that it is operated under appropriate conditions in a manner pursuant to this instruction man...

Страница 8: ...nel with the option of locking the respective electrical functions of the delivery table Control panel in the side rail Wired controller Foot controllers option 9 Backlight lamp from underneath the de...

Страница 9: ...related to the structure technology and materials without compromising the technical utility parameters and safety of the device 1 7 Safety The device structure ensures safe operation and use provide...

Страница 10: ...nvironment is described in which the medical device Delivery Table LM 03 should be operated The user should observe the guidelines and warnings included in the tables NOTE The use of accessories addit...

Страница 11: ...ransport method and package type ensuring safe transport 2 2 Unpacking and first start up NOTE If the device is mounted by the service team of FAMED YWIEC Sp z o o then the user is exempt from perform...

Страница 12: ...r the electronic table control system works performing all the functional table movements by use of all the available control elements foot controller and side panels Furthermore check whether the loc...

Страница 13: ...elected functions so as to prevent the patient from their activation with the wired controller or the ones in the side rails The central control panel enables the control of the following delivery tab...

Страница 14: ...at the same time inclination angle of the backrest and thigh sections AUTOCONTOUR Trendelenburg and reverse Trendelenburg angles The table controller without the central control panel or locking pane...

Страница 15: ...with control panels on the outer side see Fig 2a whereas on the inner side with control panels see Fig 2b which enable the activation of the table functions by the patient and medical staff as describ...

Страница 16: ...e Trendelenburg when you lower the lying surface to the min height or during the Trendelenburg position marking compliant with pt 8 2 3 3 Backrest section angular position adjustment with the CPR func...

Страница 17: ...central panel are marked with the symbol Angular couch position can be changed only by medical personnel The Trendelenburg position is a rescue position NOTE Do not lay any objects on the plastic base...

Страница 18: ...e patient s head and push the backrest section down So as to level the thigh section pull the lever under the shank section and push the thigh section down 3 10 Position 0 The position 0 of the delive...

Страница 19: ...e between the mounting element of the hand holder and the side rail Failure to observe the above note may at the extreme angles lead to crushing the patient s operator s fingers or to damaging the del...

Страница 20: ...r underneath must be free of obstacles During the table movement avoid collisions Do not roll the delivery table across electrical cables 3 15 Use of the side rails Fig 4 The side rails fig 4 are atta...

Страница 21: ...ht be damaged Application of excessive force might result in damaging the mouldings 3 17 Infusion rod The infusion rod is embedded in bushings located in the permanent lying surface frame near the hea...

Страница 22: ...ually with the agents recommended in annex 2 to this instruction manual NOTE Use foil underlays for spots exposed to preparations applied during various procedures The contact with preparations for di...

Страница 23: ...Instruction Manual for the Delivery Table FREYA 3 Page 23 4 Utility parameters The schematic figure below presents the most important utility functions of the delivery table and their marking...

Страница 24: ...tion contained herein do not operate it Consult with the manufacturer supplier dealer Using the defective device may lead to damage sustained by the user for which the manufacturer assumes no liabilit...

Страница 25: ...rotection NOTE The device contains products which may pose a threat to the environment gel battery type EP 1 2 12 2 pcs Rules of conduct as regards worn products that may pose a threat to the environm...

Страница 26: ...should be settled acc to pt 6 6 Troubleshooting So as to ensure safe and reliable functioning within the device lifetime the user shall perform inspections as regards the compliance with standard IEC...

Страница 27: ...pply cable disconnected controller defective control system or actuator connect the controller notify an authorised service centre seat section angle adjustment actuator will not work disconnected pow...

Страница 28: ...ng features The user is to ensure that the medical device Delivery Table LM 03 is used in this type of environment Resistance IEC 60601 1 2 Test level Resistance level Explanations and guidelines Elec...

Страница 29: ...a should be lower than the column resistance level within the entire frequency range b There might be interference in the medical device functioning if the column is being used in the vicinity of equ...

Страница 30: ...ective identification plate areas 1 Logo symbol manufacturer production date manufacturer name and address 2 Device symbol 3 General marking CE marking Warning Follow the instructions for safe use Not...

Страница 31: ...t 215 kg 39 st 250 kg 6 Maximum load of the extended lying surface frame 27 st 170 kg 7 Note For cleaning washing and disinfecting do not use agents which contain active oxygen or chlorine 8 Infusion...

Страница 32: ...st inclination adjustment pedal option 18 Seat inclination adjustment pedal option 19 Warning Follow the instructions for safe use 20 Follow the instructions Note The manufacturer reserves the right t...

Страница 33: ...00 3 Famed ywiec Sp z o o 6 Extension cover 1 C220920000003 B00 3 Famed ywiec Sp z o o 7 Lying surface mattress 1 C220920100000 A00 3 Famed ywiec Sp z o o 8 Extension mattress 1 C220920110000 3 Famed...

Отзывы: