FAIRCHILD MODEL HPD

Pressure Regulator

Installation, Operation and Maintenance Instructions

Panel Mounting

To panel mount the HPD, the knob must first be removed.  To remove the knob, first using a small 

screwdriver, pry up the insert in the top to expose the range screw and jam nut.  Remove the jam nut and 

then unscrew the knob from the range screw.  Insert the HPD pressure regulator through the 1 3/8" hole in 

the panel and fix it in place with the retaining nut.  Reinstall the knob on the range screw and fix it in place 

with the jam nut.  Reinstall the knob cover

Knob adjustment

The knob on the HPD pressure regulator can be adjusted to limit the range of the output pressure.  First 

assure the pressure regulator has the intended operating supply pressure applied to the unit.  Remove the 

cover from the top of the knob and loosen the jam nut on the range screw.  Insert a 3/16" hex wrench in the 

range screw and adjust the pressure regulator to the desired maximum range.  (Note, do not set the 

pressure regulator to more than 110% of the rated range.)  Adjust the knob so that it stops against the 

bonnet.  Tighten the jam nut to lock the knob to the range screw.  Reinstall the knob cover.