![FagronLab PRO Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/fagronlab/pro/pro_user-manual_538548019.webp)
FagronLab™ PRO – User Manual | 19
FagronLab™ PRO
•
DAB / DAC
•
NRF
•
Custom programs
Additional options are available
To select the program press the communicating program button. The lift arm moves into the mounting position.
The integrated camera will be activated and displayed on the screen. Scan the QR-code on the left side of the jar label, the
device will recognize the jar size.
By pressing the button "jar size", the jar size can be entered manually by selecting the communicating jar size.
Following, the mixing unit is mounted into the lift arm. The visual description shows the easy mounting of the jar into the
quick-lock coupling. By selecting "start" on the lower right, the lift arm moves automatically into the start position, the stirrer
shaft locks and the mixing process starts.
After the mixing process is finished the free spin program starts, the stirrer will be mostly clean, the mixing process is
finished and the lift arm will move to the removal position. The mixing blade unlocks and the mixing unit can be removed.
•
During the mixing process the progress of the mixing
program is displayed:
•
Following parameters are displayed:
1.
Mixing program
2.
Jar size
3.
Time (passed and remaining)
4.
List of mixing time and mixing level
5.
"Stop" button to abort the mixing process
•
A summary of the mixing process is displayed
•
It shows the used mixing program (1), the jar size (2), the
ID-Number (3) and a precise breakdown of the mixing
levels and mixing time (4) and their sum (5)
•
The identification number (ID-Number, review chapter
5) helps for an easier documentation and reproducible
product quality. We recommend to record the ID-
Number on the manufacturing record
Pic.: Display mixing progress
Pic.: Display summary mixing process
Mixing program: Pre-grinding
00:13
03:37
1000 RPM
Jar size: 30ml
01:00 min
1000 RPM
02:50 min
1200 RPM
Stop
Discard
Save
Mixing program: Pre-grinding
3 L3 d4 00 00 00 00
1
2
Σ
1000
1200
60
170
230
Jar size: 30ml
1
1
5
5
2
2
4
4
3
3
Содержание PRO