background image

FagronLab™ PRO DMS – User Manual  |  5

4. Safety instructions

Always wear protective gloves, clothes, and glasses to avoid the risk of:

 

Splashing and evaporation of liquids.

  Release of toxic or combustible gases.

Place the device in a spacious and on a stable, clean, non slip, dry and fireproof surface. Do not use the device in explosive 
atmospheres, with hazardous substances or under water.
 
Gradually increase the speed, reduce the speed if:

  Splashing, due to high speed of the stirring bar.

  The device is not running smoothly, or container moves on the base plate.

Temperature must always be set to at least 50°C lower than the ignition point.
 
Be aware of hazards due to:

  Flammable materials or utensil with a low boiling temperature.

  Overfilling of utensil.

  Unsafe container.

Process pathogenic materials only in closed vessels.

In case the stirrer bar is PTFE, please note:

  Elemental fluorine, three fluoride and alkali metals will corrode the PTFE and halogen alkanes make it expansion at room 

temperature.

  Molten alkali, alkaline earth metals or their solution, as well as the powder in second and third ethnic of the Periodic 

Table of Elements will have chemical reaction with PTFE when temperature reaches 300 - 400°C.

  Check the device and accessories prior to each use. Do not use damaged components. Safe operation is only guaranteed 

with the accessories described in the “Accessories” chapter. Accessories must be securely attached to the device and can 
not come off by themselves. Always disconnect the plug before fitting accessories.

  When the external temperature sensor needed, the tip of the measuring sensor must be at least 5-10mm from vessel 

bottom and wall.

  The device can only be disconnected from the main power supply by pulling out the main or the connector plug.

  The voltage stated on the label must correspond to the main power supply.

  Ensure that the main power supply cable does not touch the hotplate. Do not cover the device.

  The device can only be disassembled by experts.

Keep away from high magnetic field.

Warning!

  Read the operating instructions carefully before use.

  Ensure that only trained staff works with the device.

Risk of burn!

  Caution when touch the housing parts and the hotplate which can reach temperature of 340°C.

  Attention on the remaining high temperature after switching off.

Protective ground contact!

  Make sure that the power supply is grounded before use.

FagronLab™ PRO-DMS

Содержание FagronLab PRO-DMS

Страница 1: ...FagronLab PRO DMS User Manual www fagronlab com ...

Страница 2: ......

Страница 3: ...l and custom use 6 8 Control 7 Control elements 7 Control panel 7 Display 8 9 Function 8 Heating 8 Working with external temperature sensor 9 Residual heat warning HOT 9 10 Stirring 9 11 Remote control 10 12 Faults 10 13 Maintenance and cleaning 11 14 Associated standards and regulations 11 15 Specifications 12 Table of contents ...

Страница 4: ...ive with the following information Serial number on the rear panel Certification Description of problem Methods and procedures adopted to resolve the problems Your contact information This device is under warranty and free from defects in materials and workmanship under normal use and service for a period of 24 months from the date of invoice The warranty is extended only to the original purchaser...

Страница 5: ...of the Periodic Table of Elements will have chemical reaction with PTFE when temperature reaches 300 400 C Check the device and accessories prior to each use Do not use damaged components Safe operation is only guaranteed with the accessories described in the Accessories chapter Accessories must be securely attached to the device and can not come off by themselves Always disconnect the plug before...

Страница 6: ...e that the required operating voltage matches with the power supply voltage Ensure that the socket is properly grounded Plug in the power cable ensure the power is on and begin initializing Add the medium into the vessel with an appropriate stirring bar Place vessel on the work plate Set the desirable stirring speed Set the desirable temperature Stop both heating and stirring functions If these op...

Страница 7: ...b stir Set the rated rotary speed The function stirring is switched ON or OFF via pushing ON OFF the knob Heating knob heat Set the rated temperature The function heating is switched ON or OFF via pushing ON OFF the knob LCD display LCD displays the real working state and all settings LED heat When the heating function is switched ON the LED Heat is lit LED stir When the stirring function is switc...

Страница 8: ...ature of the sample 9 Function Figure 4 Device display Note If both heating and stirring functions have been started at the same time heating function always has higher priority If in this case speed is changed via the stirring knob it displays stirring speed and reverses to temperature in the duration of 5 seconds Characters Description Temp and C Display temperature when the heating function is ...

Страница 9: ...tically switch off If users want to turn off the LCD immediately just unplug the power cable When the power cable is pulled out the residual heat warning function cannot be visual 10 Stirring The function stirring is switched on or off via pushing the stirring knob The speed can be set by the knob 100 to 1500 rpm in steps of 10 rpm When both of functions heating and stirring are switched on the LC...

Страница 10: ...contact supplier The unit can be controlled from an external PC using the dedicated software via the RS232C serial interface fitted to the unit Data communication from laboratory device to computer is only possible on demand of the computer The functions of the interface lines between laboratory device and automation system are selected from the specified signals of the EIA standard RS232C corresp...

Страница 11: ... and place it in a stable dry room temperature and clean location 13 Maintenance and cleaning Proper maintenance can keep the device working properly and lengthen its lifetime Do not spray cleanser into the device when cleaning Unplug the power cable when cleaning Only use recommended cleansers to maximize the life of your FagronLab PRO DMS device Before using other method for cleaning or decontam...

Страница 12: ...ry speed display LCD Plate material Stainless steel Ceramic coated Dimensions of workplate mm Ø 135 Heating power W 500 Temperature range C RT 340 increment 1 Temperature display C LCD Temperature display accuracy C 0 1 The safe temperature of hotplate C 360 Temperature sensor in medium PT1000 Control accuracy of heating temp with temp sensor C 0 2 Residual heat warning 50 C Device Dimensions LxWx...

Страница 13: ...FagronLab PRO DMS Notes ...

Страница 14: ...FagronLab PRO DMS Notes ...

Страница 15: ......

Страница 16: ...Together we create the future of personalized medicine Fagron Hellas 12 km N R Trikala Larisa P C 42100 P O Box 32 Trikala Greece T 30 24310 83633 5 F 30 24310 83615 www fagron gr P R OD M SU M 0 1 ...

Отзывы: