SEPARIX-C/-T and SEPARIX-Control CT
Page 2/37
•
The sensors are suitable for installation in Ex Zone 0. Observe all applicable rules
and regulations for potentially explosive atmospheres.
•
The measuring transducer must be installed outside the Ex Zone.
•
The warning system must be tested at the intervals specified by local regulations.
•
To avoid the risk of electrostatic charge, the SEPARIX-C L Plus interface sensors are
not permitted to be used in heavy-flowing, non-conducting fluids (e.g. in pipework
or stirred tanks).
•
Also for cleaning the SEPARIX-C L Plus interface sensors the risk of electrostatic
charge must be prevented by the using damp cloths.
•
The SEPARIX-C L Plus interface sensor and the SEPARIX-T L Plus high-level sensor
are not permitted to be used in aggressive media.
The safety instructions in this user guide are marked as follows:
If these safety instructions are not observed, it may result in the risk of ac-
cident or damage to the SEPARIX system.
Useful tips and information in this user guide you should observe, appear
in italics and are identified by this symbol.