26
-
8-1
Operation test for residual current device
Plug the main connector and push the test button of the
device with thin stick while the mains switch is turned
on.
The condition is normal if the device works and the
mains switch is turned off.
※
If correct operation is not achieved, stop
immediately and contact the dealer or the
closest service center. Accident such as earth
leakage may occur.
CAUTION
8-2
Cleaning and caring the product
WARNING
WARNING
WARNING
CAUTION
Test button
Maintenance
・
checkup
8
If residual current device does not work properly,
electric shock hazard occurs when a short is
caused. So please conduct the test more than once
in a month.
Conduct operation test for residual
current device.
Do not disassemble the unit.
Some parts in the unit are under electric
pressure and high temperature. So disassembling
the unit may cause electric shock or cause
users physical injury.
Use appropriate product for
cleaning and caring the product in
proper way.
When cleaning and caring the product,
do not pour water directly on the external and
internal part of the unit, and also do not use
cleanser, thinner, petrol, lamp oil, acid and
related products. These products may cause
Electric shock or damage the unit.
Do not touch cooling fin with bare hands when
conducting maintenance work.
Edgy fin may cut your hands.
Do not touch cooling fin with bare
hands.
When cleaning and caring the product, turn off
the power switch and residual current breaker
and unplug the mains connector from outlet
for preventing electric shock or damage on
the product.
Unplug the mains connector when
cleaning and caring the product.