5-3 Operation in the user setting mode
In the user setting mode, the upper and lower limit temperatures for the upper and lower limit alarm [AOH]
and [AOL] and the power failure recovery function can be set.
(1) Keep pressing the [Set] key in the temperature measurement
screen (the current temperature display ON).
(2) Keep pressing the key for about 3 seconds, the display changes
to the user setting item screen, ([A-4] lamp ON)
(3) Press the [
▲
]and [
▼]
keys to select the item:
[A-4 ] Sets the power failure recovery function
[AOH] Sets the temperature for the upper limit alarm
[AOL] Sets the temperature for the lower limit alarm.
*
Transfer to the temperature measuring screen
Keep pressing the [Set] key for about 2 seconds in the user
setting mode, the display changes to the temperature setting
screen. (The current temperature display ON)
*
Transfer to the user setting mode
(1) Temperature setting screen
(2) User setting item screen
Keep pressing for
about 3 seconds
5-3-1.
Setting the power failure recovery function
(1) When selected, [A-4] and press the [Set], the parameter
is shown just before power failure. ( [A-4] is lighted.)
[on] … Power recovery function, ON ( The alarm is displayed.)
[cnt] … Power recovery function, ON ( The alarm is not displayed.)
[oFF]…Power recovery function, OFF ( The alarm is displayed.)
[dIs] … Power recovery function, OFF ( The alarm is not displayed.)
※
The setting was [on] when delivered.
※
Only when it was [on] or [off], the power recovery Alarm is shown.
(2) With the keys, [
▲
],[
▼
], select the Power recovery function setting.
(setting position was blinked.
(3) When pressing the key, [Set], the display shows the user set
parameters and the setting of the Power recovery function was
finished. ( [A-4] is lighted. )
By setting the Power failure recovery function, you can select the re-start condition whether the temperature
control and the pump start after power recovers or they do not re-start even after the power recovers.
(1) User setting item screen
(2) Power failure recovery function
(ON/OFF) selection
A-4
setting
Function
Power recovery
Alarm
on
When Power recovers, the temperature control and the pump re-start with the
parameters that were set up just before power failure.
Displayed
cnt
When Power recovers, the temperature control and the pump re-start with the
parameters that were set up just before power failure.
Not Displayed
oFF
When Power recovers, the temperature control and the pump stop no matter what the
parameters were set just before power failure.
Displayed
dIs
When Power recovers, the temperature control and the pump stop no matter what the
parameters were set just before power failure.
Not Displayed