THS30 Series Multifunction Temperature & Humidity Transmitter Operation Manual
PAGE
:
39 OF 65
Version
:
03
Date
:
2017.08.25
b.
Put the product in the humidity control box, and adjust upper point of humidity (ex: RH 80%)
c.
Wait the humidity of control box is becoming stable.
d.
Click Relative Humidity > UPPER POINT