2. Recall the factory setting of more points of humidity 

a. Click  Calibration

 

b. Click 

Relative Humidity > RECALL 

 

 

e.

 

That restore humidity of factory setting is done. 

 

 

 

 

 

 

 

 

7. Inspec tion and mainte nance  

        1. Maintenance 

           

Since this product is inspected and calibrated for high accuracy at the factory before 

a. 

b. 

Содержание THS301

Страница 1: ...THS30X Series Multifunction Temperature Humidity Transmitter Operation Manual...

Страница 2: ...ty Transmitter Operation Manual PAGE 2 OF 65 Version 03 Date 2017 08 25 1 Features 2 Security considerations 3 Installation 4 Connection 5 Display and LED instruction 6 Software setting and calibratio...

Страница 3: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 3 OF 65 Version 03 Date 2017 08 25 Oder Code...

Страница 4: ...building factory clean room Lab Monitoring for HVAC process air conditioning environmental ventilation control Temperature and humidity measuring in hospital pharmaceutical industry textile industry...

Страница 5: ...This product must be operated under the operating condition specified in this manual to prevent equipment damages This product must be operated under the normally atmospheric condition to prevent equi...

Страница 6: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 6 OF 65 Version 03 Date 2017 08 25 3 Installation 3 1 Dimension...

Страница 7: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 7 OF 65 Version 03 Date 2017 08 25...

Страница 8: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 8 OF 65 Version 03 Date 2017 08 25 3 2 Electric Connector Dimension...

Страница 9: ...30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 9 OF 65 Version 03 Date 2017 08 25 3 3 Installation BASE Installing Flange Installing Fitting thread COVER Installing Shi...

Страница 10: ...3 Date 2017 08 25 3 3 Option accessory 1 Fixed type Order code Name Desciption Dimension 4415000022 Metal Flange Material Stainless Steel 304 HS S Din rail board Material Stainless Steel 304 443500000...

Страница 11: ...40 m OD 12mm L 34 5mm Excellent filterable and pollution corrosion and pressure resistance Max Temper 200 max 8203104011 Metal grid filter with mesh Material SUS304 OD 12mm L 32mm Common resisting po...

Страница 12: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 12 OF 65 Version 03 Date 2017 08 25 4 Connection 4 1 Analogue Diagram 4 2 RS 485 Diagram 4 3 Analogue RS 485 Diagram...

Страница 13: ...ure Humidity Transmitter Operation Manual PAGE 13 OF 65 Version 03 Date 2017 08 25 5 Display and LED instruction 5 1 LCM display 5 2 LED Indication 1 Power LED P Green light ON 2 RS 485 LED P Red ligh...

Страница 14: ...S 485 ModBus Protocol 24 6 5 Display and save data 26 6 6 Choose parameter of Output 30 6 7 Temperature Calibration with two points 33 6 8 Humidity Calibration with two points 38 6 9 Temperature Calib...

Страница 15: ...ter Operation Manual PAGE 15 OF 65 Version 03 Date 2017 08 25 6 1 Application Program statement 1 Free installation program THS UI_F19 exe a Operating System requirements above Windows XP 2 Other appl...

Страница 16: ...Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 16 OF 65 Version 03 Date 2017 08 25 6 2 Setting RS 485connection 1 Connect product to PC via RS 485 cable 2 Execute THS 3 C...

Страница 17: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 17 OF 65 Version 03 Date 2017 08 25 4 Select the corresponding values of com port as fallowing 5 Click Apply a d e b c...

Страница 18: ...on Temperature Humidity Transmitter Operation Manual PAGE 18 OF 65 Version 03 Date 2017 08 25 6 Connect successfully a Show value and trend chart of Temperature and Relative Humidity b Show Open port...

Страница 19: ...smitter Operation Manual PAGE 19 OF 65 Version 03 Date 2017 08 25 6 3 Scan RS 485 connection Use scan function to connect when forgetting the connection information or having more facilites 1 Connect...

Страница 20: ...ction Temperature Humidity Transmitter Operation Manual PAGE 20 OF 65 Version 03 Date 2017 08 25 4 Select the corresponding values of com port as fallowing a Port b RS 485 4 Click Scan to execute conn...

Страница 21: ...30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 21 OF 65 Version 03 Date 2017 08 25 5 Scan connection facilities and set up a Select Station ID b Click CLOSE AND EXPORT...

Страница 22: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 22 OF 65 Version 03 Date 2017 08 25 6 Click Apply...

Страница 23: ...tion Temperature Humidity Transmitter Operation Manual PAGE 23 OF 65 Version 03 Date 2017 08 25 7 Connect successfully a Show values and trend chat Temperature and Relative Humidity b Show Open port R...

Страница 24: ...eries Multifunction Temperature Humidity Transmitter Operation Manual PAGE 24 OF 65 Version 03 Date 2017 08 25 6 4 Setting RS 485 ModBus Protocol 1 Setting RS 485 connection step as step 6 1 2 Click S...

Страница 25: ...rsion 03 Date 2017 08 25 3 Select Modbus Protocol parameter a Station ID 1 247 b Baud Rate 9600 19200 38400 57600 115200 c Data Frame None 8Bit 1Stop None 8Bit 2Stop Even 8Bit 1Stop Even 8Bit 2Stop Od...

Страница 26: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 26 OF 65 Version 03 Date 2017 08 25 6 5 Display and save data 1 Show data Click Display...

Страница 27: ...yle Select the OUTPUT channel Set line color chosen from OUTPUT Snap chart Save the data measuring when the system start connecting before clinking the Export icon Show Not show the statement of Y axi...

Страница 28: ...S30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 28 OF 65 Version 03 Date 2017 08 25 3 Setting time interval of record a File Log Interval b Select time interval of reco...

Страница 29: ...03 Date 2017 08 25 4 Save Log measuring data a Log measuring range Save the data measuring when the system start connecting before clinking the Export icon a 1 Click Display Export a 2 Appoint path a...

Страница 30: ...F 65 Version 03 Date 2017 08 25 b Log measuring data Log the data which is on from start or off b 1 Display Log OFF b 1 Appoint path and Key in file name save Log ON 1 If file name is some as the path...

Страница 31: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 31 OF 65 Version 03 Date 2017 08 25...

Страница 32: ...eration Manual PAGE 32 OF 65 Version 03 Date 2017 08 25 2 Select relative parameters from Output1and Output2 a Output style b Responding time c Voltage or currant Output d Voltage or current analog ra...

Страница 33: ...es Multifunction Temperature Humidity Transmitter Operation Manual PAGE 33 OF 65 Version 03 Date 2017 08 25 6 7 Temperature Calibration with two points 1 Calibrate upper point of temperature a Click C...

Страница 34: ...smitter Operation Manual PAGE 34 OF 65 Version 03 Date 2017 08 25 b Put the product in the temperature control box and adjust the upper point of temperature ex 100 c Wait the temperature of control bo...

Страница 35: ...ries Multifunction Temperature Humidity Transmitter Operation Manual PAGE 35 OF 65 Version 03 Date 2017 08 25 e Input the reference temperarue then click SET f That calibrating upper point temperature...

Страница 36: ...7 08 25 2 Calibrate lower point of temperature a Put the product in the temperature control box and adjust the lower point of temperature ex 0 b The difference temperature between Upper and Lower poin...

Страница 37: ...eries Multifunction Temperature Humidity Transmitter Operation Manual PAGE 37 OF 65 Version 03 Date 2017 08 25 e Input reference temperature then click SET f That calibrating lower point temperature i...

Страница 38: ...eries Multifunction Temperature Humidity Transmitter Operation Manual PAGE 38 OF 65 Version 03 Date 2017 08 25 6 8 Humidity Calibration with two points 1 Calibrate Upper point of humidity a Click Cali...

Страница 39: ...ansmitter Operation Manual PAGE 39 OF 65 Version 03 Date 2017 08 25 b Put the product in the humidity control box and adjust upper point of humidity ex RH 80 c Wait the humidity of control box is beco...

Страница 40: ...0 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 40 OF 65 Version 03 Date 2017 08 25 e Input reference humidity then click SET f That calibrating upper point humidity is d...

Страница 41: ...ion Manual PAGE 41 OF 65 Version 03 Date 2017 08 25 2 Calibrate humidity lower point a Put the product in the humidity control box and adjust lower point of humidity ex RH 20 b Wait the humidity of co...

Страница 42: ...0 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 42 OF 65 Version 03 Date 2017 08 25 d Input reference humidity then click SET e That calibrating lower point humidity is d...

Страница 43: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 43 OF 65 Version 03 Date 2017 08 25 6 9 Temperature Calibration with signal points 1 Click Calibration...

Страница 44: ...2017 08 25 2 Put the product in the temperature control box and adjust the temperature point which you want to calibrate ex 50 3 Wait the temperature of control box is becoming stable 4 Click Temperat...

Страница 45: ...eries Multifunction Temperature Humidity Transmitter Operation Manual PAGE 45 OF 65 Version 03 Date 2017 08 25 5 Input reference temperature then click SET 6 That calibrating signal point temperature...

Страница 46: ...THS30 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 46 OF 65 Version 03 Date 2017 08 25 6 10 Humidity Calibration with signal point 1 Click Calibration...

Страница 47: ...Operation Manual PAGE 47 OF 65 Version 03 Date 2017 08 25 2 Put the product in humidity control box then adjust the humidity point which you want to calibrate ex 50 3 Wait the humidity of control box...

Страница 48: ...0 Series Multifunction Temperature Humidity Transmitter Operation Manual PAGE 48 OF 65 Version 03 Date 2017 08 25 5 Input reference humidity then click SET 6 That calibrating signal point humidity is...

Страница 49: ...E 49 OF 65 Version 03 Date 2017 08 25 6 11 Restore factory setting of signal two point s 1 Restore factory setting temperature a Click Calibration b Click temperature RECALL FACTORY CALIB c Show store...

Страница 50: ...er Operation Manual PAGE 50 OF 65 Version 03 Date 2017 08 25 2 Restore humidity of factory setting a Click Calibration b Clock Relative Humidity RECALL FACTORY CALIB c Show store Factory Rel Humidity...

Страница 51: ...ause of that the product has high accuracy and using this calibration way can influence linear accuracy we do not suggest to use this calibration way 1 click Calibration Statement The existing value i...

Страница 52: ...e 4 Retain the factory setting a Input the value which you want to calibrate in the Temperature area ex 25 a 1 Tunit Value which product shows a 2 Tstd Standard calibration value 1 Please enter calibr...

Страница 53: ...tifunction Temperature Humidity Transmitter Operation Manual PAGE 53 OF 65 Version 03 Date 2017 08 25 5 Delete factory setting a click Temperature CLEAR clear data b Click Temperature WRITE clear fact...

Страница 54: ...rsion 03 Date 2017 08 25 c Input the value which you want to calibrate c 1 Tunit Value which product shows c 2 Tstd Standard value of calibration 1 Please enter calibration points in the blank 10 poin...

Страница 55: ...OF 65 Version 03 Date 2017 08 25 6 Retain the part of factory setting a Click left keyboard twice on the mouse on the factory setting which you want to delete b Click delete icon of Keyboard or right...

Страница 56: ...ture Humidity Transmitter Operation Manual PAGE 56 OF 65 Version 03 Date 2017 08 25 d Input the value which you want to calibrate d 1 Tunit Value which product shows d 2 Tstd Standard value of calibra...

Страница 57: ...65 Version 03 Date 2017 08 25 6 13 Humidity Calibration with more points 1 Click Calibration Statement The existing value is record of more points of calibrating which was executed by factory when op...

Страница 58: ...e 4 Retain the factory setting a Input the value which you want to calibrate in the Relative Humidity area a 1 Hunit Value which product shows a 2 Hstd Standard value of calibration 1 Please enter cal...

Страница 59: ...tion Temperature Humidity Transmitter Operation Manual PAGE 59 OF 65 Version 03 Date 2017 08 25 5 Delete factory setting a click Relative Humidity CLEAR clear data b Click Relative Humidity WRITE clea...

Страница 60: ...08 25 c Input the value which you want to calibrate in the Relative Humidity area c 1 Hunit Value which product shows c 2 Hstd Standard value of calibration 1 Please enter calibration points in the b...

Страница 61: ...65 Version 03 Date 2017 08 25 6 Retain the part of factory setting a Click left keyboard twice on the mouse on the factory setting which you want to delete b Click delete icon of Keyboard or right key...

Страница 62: ...smitter Operation Manual PAGE 62 OF 65 Version 03 Date 2017 08 25 d Input the value which you want to calibrate in the Relative Humidity area d 1 Hunit Value which product shows d 2 Hstd Standard valu...

Страница 63: ...Operation Manual PAGE 63 OF 65 Version 03 Date 2017 08 25 6 14 Restore factory setting of more points 1 Recall temperature of factory setting a Click Calibration b Click Temperature RECALL d That res...

Страница 64: ...2017 08 25 2 Recall the factory setting of more points of humidity a Click Calibration b Click Relative Humidity RECALL e That restore humidity of factory setting is done 7 Inspection and maintenance...

Страница 65: ...ance Do not damage sensor surface during maintenance process 3 Troubleshooting If any problem occurs during operation refer to the table below for appropriate solutions 2 Troubleshooting Remove the se...