5                                          

 SDL310-EU-EN V1.4   07/13

 

Datalogger 

Types of Data Recording 

 

Manual Datalogging:

 Manually log up to 99 readings onto an SD card via push-key press. 

 

Automatic Datalogging:

 Automatically log data onto an SD memory card where the number of 

data points is limited only by the card size. Readings are logged at a rate specified by the user. 

SD Card Information 

  Insert an SD card (from 1G size up to 16G) into the SD card slot at the bottom of the meter. 

With regard to orientation, the card should be inserted with the front of the card (label side) 
facing toward the rear of the meter. 

  If the SD card is being used for the first time it is recommended that the card be formatted and 

the logger’s clock set to allow for accurate date/time stamping during datalogging sessions. 
Refer to the Setup Mode section for SD card formatting and time/date setting instructions. 

  European and USA numerical formats differ. The data on the SD card can be formatted for 

either format. The meter defaults to USA mode where a decimal point is used to separate units 
from tenths, i.e. 

20.00

. The European format uses a comma, i.e. 

20,00

. To change this setting, 

refer to the Setup Mode section. 

Manual Datalogging 

In the manual mode the user presses the LOG key to manually log a reading onto the SD card. 
1. 

Set the sampling rate to ‘0’ seconds. 

2. 

Press and hold the LOG key for at least 1.5 seconds; the lower portion of the display will show 
p-

n

 (

n

 = memory position number 1-99). 

3. 

Press the LOG key momentarily to log a reading into memory. The REC icon will flash each 
time a data point is stored (the SCAN SD icon will appear when the meter accesses the card). 

4. Use 

the 

 and 

 keys to select one of the 99 data memory positions in which to record.   

5. 

To exit the manual datalogging mode, press and hold the LOG key for at least 1.5 seconds. 

Automatic Datalogging 

In automatic datalogging mode the meter takes and stores a reading at a user-specified sampling 
rate onto SD card. The meter defaults to a sampling rate of one second. To change the sampling 
rate, refer to the Setup Mode section (the sampling rate cannot be ‘0’ for automatic datalogging): 
1. 

To begin an automatic Datalog session press and hold the LOG key for at least 1.5 seconds. 

2. 

The meter will scan for an SD card and verify that it can be used to store data. If a card is not 
inserted or if the card is defective, the meter will display SCAN SD indefinitely. In this case, 
switch the meter OFF and try again with a valid SD card. 

3. 

If the SD card is valid, the display will show the LOG icon (or the LOG icon alternating with the 
temperature display) and then the REC icon will flash each time that a reading is stored. 

4. 

To pause the datalogger, press the LOG key momentarily. The REC icon will stop flashing. To 
resume logging simply press the LOG key again momentarily. 

5. 

To terminate the datalogging session press and hold the LOG key for at least 1.5 seconds. 

6. 

When an SD card is used for the first time a folder is created on the card and named 

AMC01

Up to 99 spreadsheet documents (each with 30,000 readings) can be stored in this folder. 

7. 

When datalogging begins a new spreadsheet document named 

AMC01001.xls

 is created on 

the SD card in the AMC01 folder. The data recorded will be placed in the AMC01001.xls 
document until 30,000 readings are reached. 

8. 

If the measurement session exceeds 30,000 readings, a new document will be created 
(AMC01002.xls) where another 30,000 readings can be stored. This method continues for up 
to 99 documents, after which another folder is created (AMC02) where another 99 spreadsheet 
documents can be stored. This process continues in this same fashion with folders AMC03 
through AMC10 (last allowable folder). 

   www.ponpe.com

contact:

[email protected]

Содержание SDL310

Страница 1: ...USER GUIDE Vane Thermo Anemometer Datalogger Model SDL310 www ponpe com contact sale ponpe com ...

Страница 2: ...de Extech Instruments is an ISO 9001 certified company Safety International Safety Symbols This symbol adjacent to another symbol or terminal indicates the user must refer to the manual for further information Meter Description 1 Thermocouple input 2 Vane input plug 3 Air velocity reading 4 Units of measure 5 HOLD Backlight key 6 MAX MIN key 7 SET and Clock key 8 PC interface jack 9 Reset key 10 P...

Страница 3: ... minute Km h kilometers per hour knots and mph miles per hour 1 Press and hold the UNIT key for at least 1 5 seconds to switch from one unit of measure to another 2 Note that pressing and holding the UNIT key continuously allows for quicker scrolling Release the key when the desired unit of measure is displayed Thermcouple Type J or K Temperature Mode 1 Select the Thermocouple Temperature mode usi...

Страница 4: ...light ON or OFF press and hold the backlight key for at least 1 5 seconds The meter will beep when switching the backlight ON or OFF unless the beeper is disabled as described in the Setup Mode section of this user guide System Reset If the meter s keys become inoperable or if the display freezes the Reset button can be used to reset the instrument 1 Use a paper clip or similar item to momentarily...

Страница 5: ...ess and hold the LOG key for at least 1 5 seconds Automatic Datalogging In automatic datalogging mode the meter takes and stores a reading at a user specified sampling rate onto SD card The meter defaults to a sampling rate of one second To change the sampling rate refer to the Setup Mode section the sampling rate cannot be 0 for automatic datalogging 1 To begin an automatic Datalog session press ...

Страница 6: ...lable at most outlets where computer accessories are sold 4 Power the PC and run a spreadsheet software program Open the saved documents in the spreadsheet software program see examples of spreadsheet displays below Spreadsheet data example Spreadsheet example Plotting the data RS 232 USB PC Interface For streaming of data to a PC via the RS232 Output jack the optional 407001 USB kit RS232 to USB ...

Страница 7: ...ard Format t CF Select the Temperature unit of measure to C or F tYPE Select the thermocouple type to K or J Setting the Clock Time 1 Access the dAtE parameter 2 Use the arrow keys to change a value 3 Use the ENTER button to step through the selections 4 Press and hold the SET button for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatical...

Страница 8: ...operating mode SD Card FORMATTING 1 Access the Sd F parameter 2 Use the arrow keys to select YES to format the card select NO to abort Note that all data on the card will be lost if formatting is attempted 3 Press ENTER to confirm selection 4 Press ENTER again to re confirm 5 The meter will automatically return to the normal operating mode when formatting is complete If not press and hold the SET ...

Страница 9: ...lost 3 Replace the six 6 1 5V AA batteries observing polarity 4 Replace the battery compartment cover with the two 2 Phillips screws All EU users are legally bound by the Battery Ordinance to return all used batteries to community collection points or wherever batteries accumulators are sold Disposal in household trash or refuse is prohibited Disposal Follow the valid legal stipulations in respect...

Страница 10: ...tion Automatic temperature compensation for the Anemometer function and the type K J thermometer function Data Hold Freeze the display reading Memory Recall Record and Recall the Maximum and Minimum readings Display update rate Approx 1 second Data Output RS 232 USB PC computer interface Operating Temperature 0 to 50 C 32 to 122 F Operating Humidity 85 R H max Auto Power OFF After 10 minutes of in...

Страница 11: ... Type K J Thermocouple Thermometer Temperature Sensor Type Resolution Range Accuracy 0 1 C 50 0 to 1300 0 C 100 0 to 50 1C 0 4 rdg 0 5 C 0 4 rdg 1 C Type K 0 1 F 58 0 to 2372 0 F 148 0 to 58 1 F 0 4 rdg 1 F 0 4 rdg 1 8 F 0 1 C 50 0 to 1200 0 C 100 0 to 50 1 C 0 4 rdg 0 5 C 0 4 rdg 1 C Type J 0 1 F 58 0 to 2192 0 F 148 0 to 58 1 F 0 4 rdg 1 F 0 4 rdg 1 8 F Note Above specifications tested under an ...

Отзывы: