MAX-MIN Readings 

For a given measurement session, this meter can record the highest (MAX) and the lowest (MIN) 
readings for later recall. 
1. 

Press the MAX-MIN key momentarily to access this mode of operation (REC icon appears) 

2. 

The meter is now recording the MAX and MIN readings. 

3. 

Press the MAX-MIN key again to view the current MAX reading (MAX icon appears). The 
reading on the display is now the highest reading encountered since the REC icon was 
switched on (when the MAX-MIN key was first pressed). 

4. 

Press the MAX-MIN key again to view the current MIN reading (MIN icon appears). The 
reading on the display is now the lowest reading encountered since the REC icon was 
switched on (when the MAX-MIN key was first pressed). 

5. 

To exit the MAX-MIN mode, press and hold the MAX-MIN key for at least 1.5 seconds. The 
meter will beep, the REC-MAX-MIN icons will switch off, the memory will clear, and the meter 
will return to the normal operating mode.  

Display Backlight 

To turn the display backlight ON or OFF, press and hold the backlight 

 key for at least 1.5 

seconds. The meter will beep when switching the backlight ON or OFF unless the beeper is 
disabled as described in the Setup Mode section of this user guide. 

System Reset 

If the meter’s keys become inoperable or if the display freezes, the Reset button can be used to 
reset the instrument. 
1. 

Use a paper clip or similar item to momentarily press the reset button located on the lower right 
side of the instrument under the snap-off compartment cover. 

2. 

After pressing the Reset button, switch the instrument ON by pressing and holding the 
POWER key for at least 1.5 seconds. If using the power adaptor, unplug the adaptor and then 
plug it back in again to power the meter. 

AC Power Adaptor 

This meter is normally powered by six (6) 1.5V ‘AA’ batteries. However, an optional AC 9V power 
adaptor is available. To use the adaptor, insert the power adaptor plug into the jack on the bottom 
right side of the meter (under the snap-off compartment cover); plug the other end of the adaptor 
into an AC power source. The meter will now be permanently powered (as long as the adaptor is 
used) and the power key will be disabled. 
 

   www.ponpe.com

contact:

[email protected]

Содержание SDL310

Страница 1: ...USER GUIDE Vane Thermo Anemometer Datalogger Model SDL310 www ponpe com contact sale ponpe com ...

Страница 2: ...de Extech Instruments is an ISO 9001 certified company Safety International Safety Symbols This symbol adjacent to another symbol or terminal indicates the user must refer to the manual for further information Meter Description 1 Thermocouple input 2 Vane input plug 3 Air velocity reading 4 Units of measure 5 HOLD Backlight key 6 MAX MIN key 7 SET and Clock key 8 PC interface jack 9 Reset key 10 P...

Страница 3: ... minute Km h kilometers per hour knots and mph miles per hour 1 Press and hold the UNIT key for at least 1 5 seconds to switch from one unit of measure to another 2 Note that pressing and holding the UNIT key continuously allows for quicker scrolling Release the key when the desired unit of measure is displayed Thermcouple Type J or K Temperature Mode 1 Select the Thermocouple Temperature mode usi...

Страница 4: ...light ON or OFF press and hold the backlight key for at least 1 5 seconds The meter will beep when switching the backlight ON or OFF unless the beeper is disabled as described in the Setup Mode section of this user guide System Reset If the meter s keys become inoperable or if the display freezes the Reset button can be used to reset the instrument 1 Use a paper clip or similar item to momentarily...

Страница 5: ...ess and hold the LOG key for at least 1 5 seconds Automatic Datalogging In automatic datalogging mode the meter takes and stores a reading at a user specified sampling rate onto SD card The meter defaults to a sampling rate of one second To change the sampling rate refer to the Setup Mode section the sampling rate cannot be 0 for automatic datalogging 1 To begin an automatic Datalog session press ...

Страница 6: ...lable at most outlets where computer accessories are sold 4 Power the PC and run a spreadsheet software program Open the saved documents in the spreadsheet software program see examples of spreadsheet displays below Spreadsheet data example Spreadsheet example Plotting the data RS 232 USB PC Interface For streaming of data to a PC via the RS232 Output jack the optional 407001 USB kit RS232 to USB ...

Страница 7: ...ard Format t CF Select the Temperature unit of measure to C or F tYPE Select the thermocouple type to K or J Setting the Clock Time 1 Access the dAtE parameter 2 Use the arrow keys to change a value 3 Use the ENTER button to step through the selections 4 Press and hold the SET button for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatical...

Страница 8: ...operating mode SD Card FORMATTING 1 Access the Sd F parameter 2 Use the arrow keys to select YES to format the card select NO to abort Note that all data on the card will be lost if formatting is attempted 3 Press ENTER to confirm selection 4 Press ENTER again to re confirm 5 The meter will automatically return to the normal operating mode when formatting is complete If not press and hold the SET ...

Страница 9: ...lost 3 Replace the six 6 1 5V AA batteries observing polarity 4 Replace the battery compartment cover with the two 2 Phillips screws All EU users are legally bound by the Battery Ordinance to return all used batteries to community collection points or wherever batteries accumulators are sold Disposal in household trash or refuse is prohibited Disposal Follow the valid legal stipulations in respect...

Страница 10: ...tion Automatic temperature compensation for the Anemometer function and the type K J thermometer function Data Hold Freeze the display reading Memory Recall Record and Recall the Maximum and Minimum readings Display update rate Approx 1 second Data Output RS 232 USB PC computer interface Operating Temperature 0 to 50 C 32 to 122 F Operating Humidity 85 R H max Auto Power OFF After 10 minutes of in...

Страница 11: ... Type K J Thermocouple Thermometer Temperature Sensor Type Resolution Range Accuracy 0 1 C 50 0 to 1300 0 C 100 0 to 50 1C 0 4 rdg 0 5 C 0 4 rdg 1 C Type K 0 1 F 58 0 to 2372 0 F 148 0 to 58 1 F 0 4 rdg 1 F 0 4 rdg 1 8 F 0 1 C 50 0 to 1200 0 C 100 0 to 50 1 C 0 4 rdg 0 5 C 0 4 rdg 1 C Type J 0 1 F 58 0 to 2192 0 F 148 0 to 58 1 F 0 4 rdg 1 F 0 4 rdg 1 8 F Note Above specifications tested under an ...

Отзывы: