Electrode Maintenance and Troubleshooting 

Most pH measurement difficulties are caused by the electrode. Careful handling and proper 
storage of the electrode will increase measurement accuracy and electrode life. 

Electrodes should be rinsed to remove all traces of storage solution, process medium, or 
previous test solution. 

 

Electrode Storage 

The electrode should be stored in a soaker bottle or wetting cap filled with pH4 buffer. 
Never store an electrode in distilled or deionized water as this will cause migration of the fill 
solution from the electrode. 

Electrode Cleaning 

The type of electrode coating that occurs over time will determine the cleaning method. 
Soft coatings can be removed by vigorous stirring or by using a squirt bottle. Organic 
chemical coating (hard coatings) should be chemically removed. Only in extreme cases 
should the bulb be cleaned mechanically. 
 

Troubleshooting an Electrode 

Symptom

 

Cause

 

Recommendation

 

Clogged Junction 

Soak in 4.07 M Potassium Chloride (KCl) 
@ 60

o

C for 30 minutes. 

Strong Alkaline measurement 

Soak in 0.1 M HCL overnight. 

Deteriorated Gel layer 

Replace electrode. 

Protein coating on electrode 
surface 

Soak in 1 gm Pepsin dissolved in 100ml of 
0.1 m HCL for 30 minutes or as needed. 

Oil, paint, dyes, suspended 
solids etc. are on sensor 

Rinse electrode alternately with a materials 
solvent and a buffer 7.00. 

Organic solvents are coating 
the sensor 

Organic mole fraction has to be less than 
50% to assure reasonable readings. Limit 
time of measurement. Keep probe in the 
7.0 buffer between readings. 

Long response time or 
reading drift 

Dehydrated membrane 

Read solution for dry bulb below: 

Dry Bulb 

Long term storage without 
wetting  

Soak electrode tip in wetting cap filled with 
1ml 7.00 buffer for 24 to 48 hours 

Static Charge 

Wiping electrodes 

Rinse electrode in 7.0 buffer and blot. Do 
no wipe electrode. 

Same readings in different 
buffers and samples 

Cracked or broken bulb 

Replace electrode. Use bulb guard. Avoid 
plunging electrode to bottom of container 
and spinning bars. Wetting cap will protect 
bulb between tests. 

Erratic LCD display 

Samples have low ionic 
strength (lacks salt); e.g. 
distilled, de-ionized, boiled, 
lake water (high pressure) 

For each 50 ml of sample add 1 drop 
(50uL) of SAT. KCl (No alteration in pH will 
occur using inert KCl). 

Содержание PH220A

Страница 1: ...evice with tactile push buttons is battery operated and ideal for field use The PH220 simultaneously displays pH and temperature The rubber holster which protects the meter has a magnetic backing for...

Страница 2: ...simultaneous pH and temperature readings Over range or open indication Displays HHH for over range or open input conditions Displays LLL for under range pH Calibration points 4 00 7 00 10 00 1 2 or 3...

Страница 3: ...play icons are used and are discussed in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described later in this manual 2 Determine the test type pH or ORP an...

Страница 4: ...nction does not require calibration pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automa...

Страница 5: ...g the on off button Wait for the self test SLF tSt to finish 4 Use the MODE button to select mV mode The display will reflect the unit of measure 5 Place the electrode into a sample solution When the...

Страница 6: ...time the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and ho...

Страница 7: ...it over the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagra...

Страница 8: ...ine measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1 m HCL for 30 minutes or as neede...

Страница 9: ...o sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authori...

Отзывы: