Extech Instruments PH220A Скачать руководство пользователя страница 5

Model PH220 Version 3.1 11/2005 

Measuring pH 

Electrodes should always be rinsed before and between samples 
with distilled or DI (de-ionized) water to remove all traces of 
storage solution, process media, or previous test solution. 

1. 

Connect the pH electrode to the meter. 

2. 

Turn the meter on using the on/off button       . Wait for the self test (‘SLF tSt’) to 
finish. 

3. Use 

the 

MODE

 button to select pH mode. The unit of measure appears on the LCD. 

4. 

To change the unit of measure for temperature (

o

C to 

o

F or 

o

F to 

o

C), press and hold 

the 

MODE

 button for 3 seconds. The LCD reflects the current unit of measure. 

5. 

Rinse the electrode with distilled or deionized water. 

6. 

Place the electrode in the solution. Allow 30 seconds for the electrode/ATC to reach 
thermal equilibrium with the buffer solution. 

7. 

After the measurement stabilizes, view the pH reading on the main display. An out of 
range reading is indicated by the display icon ‘HHH’. Note that ‘HHH’ will also display 
when the electrode is disconnected from the meter. 

8. 

When the test is complete rinse the electrode with distilled or deionized water. 

Measuring ORP (millivolts mV) 

1. 

Connect the ORP electrode to the meter. 

2. 

Rinse the ORP electrode  

3. 

Turn the meter on using the on/off button      . Wait for the self test (‘SLF tSt’) to finish. 

4. Use 

the 

MODE

 button to select mV mode. The display will reflect the unit of measure. 

5. 

Place the electrode into a sample solution. When the LCD stabilizes it will read the 
redox potential of the sample in millivolts (mV). 

6. 

When the test is complete rinse the electrode with distilled or deionized water. 

Automatic Temperature Compensation (ATC) 

The millivolt output of a pH electrode varies with temperature, and for this reason the meter 
must compensate for these variations to obtain accurate measurements. The PH220 
automatically compensates for solution temperature changes with the electrode’s built-in 
thermometer. The temperature compensation is active when the solution temperature 
deviates from room temperature. The temperature of a solution is displayed on the lower 
LCD line of the instrument.  

Automatic Power OFF 

To conserve battery power, the instrument automatically shuts off after 10 minutes of 
inactivity. To disable or enable this feature, follow the steps below: 

 

To disable: With the power on, press and hold the on/off button       for 2 seconds 
(oFF and AtP will display). The Auto Power OFF feature is now disabled. 

 

To enable: With the power on, press and hold the on/off button for 2 seconds (on 
and AtP will display). The Auto Power OFF feature is now enabled. 

 

Each time the meter is turned off the Auto power OFF feature is set to the enabled 
mode. 

Содержание PH220A

Страница 1: ...evice with tactile push buttons is battery operated and ideal for field use The PH220 simultaneously displays pH and temperature The rubber holster which protects the meter has a magnetic backing for...

Страница 2: ...simultaneous pH and temperature readings Over range or open indication Displays HHH for over range or open input conditions Displays LLL for under range pH Calibration points 4 00 7 00 10 00 1 2 or 3...

Страница 3: ...play icons are used and are discussed in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described later in this manual 2 Determine the test type pH or ORP an...

Страница 4: ...nction does not require calibration pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automa...

Страница 5: ...g the on off button Wait for the self test SLF tSt to finish 4 Use the MODE button to select mV mode The display will reflect the unit of measure 5 Place the electrode into a sample solution When the...

Страница 6: ...time the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and ho...

Страница 7: ...it over the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagra...

Страница 8: ...ine measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1 m HCL for 30 minutes or as neede...

Страница 9: ...o sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authori...

Отзывы: