Model PH220 Version 3.1 11/2005 

Storing and Recalling Readings 

This instrument has the ability to store up to 25 measurement readings. At a later time the 
readings can be recalled and viewed. Readings are saved even after the instrument has 
been turned off. 

STORE Readings 

The steps below cover the 

STORE

 process in detail: 

1. 

Take a measurement as described earlier in this manual. 

2. 

Momentarily press the 

STORE/RECALL

 button to save the reading. The reading will 

be held on the display (the HOLD and STORE icons will appear). In addition, the 
storage location (1 to 25) will briefly appear on the lower portion of the LCD. 

3. 

Note: The temperature reading associated with the pH reading is also saved. 

4. Press 

the 

STORE/RECALL

 button again to release the ‘hold’ feature. The instrument 

will then return to normal operation. 

5. 

Note that when all 25 storage locations are filled the instrument begins writing over 
the previously stored data. 

RECALL Readings 

The steps below cover the 

RECALL

 process in detail: 

1. 

Press and hold the 

STORE/RECALL

 button until a number appears on the main 

display. This number is the storage location for the most recent reading. 

2. 

The number will switch off after 1 second and the stored reading will appear. The 
RECALL icon will appear and remain on the entire time the recall mode is accessed. 

3. 

While in the RECALL mode, use the 

STORE/RECALL

 button to scroll through the 

stored readings. The most recent readings will show first (LAST IN – FIRST OUT). 

4. 

Press and hold the 

STORE/RECALL

 button until the END display appears. Release 

the button and the instrument will return to normal operation mode. 

5. 

Note that if there are no readings in memory, the display will show END when the 
user attempts to access the RECALL mode. 

ERASE Readings 

The steps below cover the 

ERASE 

(CLEAR) process in detail: 

1. 

Press and hold 

STORE/RECALL

 and the 

CAL

 buttons simultaneously for 4 seconds. 

2. 

The ‘CLr’ display will appear indicating that all readings have been erased. 

3. 

Release the buttons to return to normal operation mode. 

Temperature Offset Adjustment 

Applications may arise where it is desirable to offset the instrument’s temperature display. 
This instrument allows a temperature display offset adjustment of ±9

o

F (±5

o

C). To do so, 

follow the steps below: 

1. 

Press and hold the 

CAL

MODE

, and 

STORE/RECALL

 buttons simultaneously for 6 

seconds. 

2. 

The main display will indicate the current offset. The lower display will show ‘t_O’. 

3. Use 

the 

CAL

 button to decrease the offset value and the 

MODE

 button to increase it. 

4. 

Press and hold the 

STORE/RECALL

 button for 2 seconds to exit this mode. The 

instrument will display ‘SA’ (save) and ‘END’ before returning to the normal operation 
mode. 

5. 

The temperature display will now indicate the actual temperature ± the offset value. 

Содержание PH220A

Страница 1: ...evice with tactile push buttons is battery operated and ideal for field use The PH220 simultaneously displays pH and temperature The rubber holster which protects the meter has a magnetic backing for...

Страница 2: ...simultaneous pH and temperature readings Over range or open indication Displays HHH for over range or open input conditions Displays LLL for under range pH Calibration points 4 00 7 00 10 00 1 2 or 3...

Страница 3: ...play icons are used and are discussed in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described later in this manual 2 Determine the test type pH or ORP an...

Страница 4: ...nction does not require calibration pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automa...

Страница 5: ...g the on off button Wait for the self test SLF tSt to finish 4 Use the MODE button to select mV mode The display will reflect the unit of measure 5 Place the electrode into a sample solution When the...

Страница 6: ...time the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and ho...

Страница 7: ...it over the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagra...

Страница 8: ...ine measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1 m HCL for 30 minutes or as neede...

Страница 9: ...o sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authori...

Отзывы: