background image

 

User Guide 

 

 

 

Waterproof Palm pH Meter 

Model PH220A 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Introduction 

Congratulations on your purchase of the Extech PH220 waterproof pH/mV meter. This 
microprocessor-based device with tactile push-buttons is battery operated and ideal for 
field use. The PH220 simultaneously displays pH and temperature. The rubber holster 
which protects the meter has a magnetic backing for convenient mounting. The built-in 
memory (25 readings) allows STORE, RECALL, and CLEAR functionality. Careful use of 
this meter will provide years of reliable service. 

7.00

Содержание PH220A

Страница 1: ...evice with tactile push buttons is battery operated and ideal for field use The PH220 simultaneously displays pH and temperature The rubber holster which protects the meter has a magnetic backing for...

Страница 2: ...simultaneous pH and temperature readings Over range or open indication Displays HHH for over range or open input conditions Displays LLL for under range pH Calibration points 4 00 7 00 10 00 1 2 or 3...

Страница 3: ...play icons are used and are discussed in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described later in this manual 2 Determine the test type pH or ORP an...

Страница 4: ...nction does not require calibration pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automa...

Страница 5: ...g the on off button Wait for the self test SLF tSt to finish 4 Use the MODE button to select mV mode The display will reflect the unit of measure 5 Place the electrode into a sample solution When the...

Страница 6: ...time the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and ho...

Страница 7: ...it over the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagra...

Страница 8: ...ine measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1 m HCL for 30 minutes or as neede...

Страница 9: ...o sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authori...

Отзывы: