background image

                                                          

8

                                                   PH220-en-US_v3.5   3/21

 

 

Electrode Maintenance and Troubleshooting 

Most pH measurement difficulties can be traced to the electrode. Careful handling and proper 
storage of the electrode will optimize measurement accuracy and electrode life. Electrodes should 
be rinsed to remove all traces of storage solution, process medium, or previous test solution. 

Electrode Storage 

The electrode should be stored in a soaker bottle or wetting cap filled with pH4 buffer. Never store 
an electrode in distilled or deionized water as this will cause migration of the fill solution from the 
electrode. 

Electrode Cleaning 

The  type  of  electrode  coating  that  occurs  over  time  will  determine  the  cleaning  method.  Soft 
coatings can be removed by vigorous stirring or by using a squirt bottle. Organic chemical coating 
(hard coatings) should be chemically removed. Only in extreme cases should the bulb be cleaned 
mechanically. 

Electrode Troubleshooting 

Symptom 

Cause 

Recommendation 

Long response time or 
reading drift 

Clogged Junction 

Soak in 4.07 M Potassium Chloride (KCl) @ 
140

o

F (60

o

C) for 30 minutes. 

Strong Alkaline measurement 

Soak in 0.1 M HCL overnight. 

Deteriorated Gel layer 

Replace electrode. 

Protein coating on electrode 
surface 

Soak in 1 gm Pepsin dissolved in 100ml of 0.1 
m HCL for 30 minutes or as needed. 
 

Oil, paint, dyes, suspended 
solids etc. are on sensor 

Rinse electrode alternately with a materials 
solvent and a buffer 7.00. 
 

Organic solvents are coating the 
sensor 

Organic mole fraction should be less than 50% 
to assure reasonable readings. Limit time of 
measurement. Keep probe in the 7.0 buffer 
between readings. 

Dehydrated membrane 

Read solution for dry bulb below: 

Dry Bulb 

Long term storage without 
wetting  

Soak electrode tip in wetting cap filled with 
1ml 7.00 buffer for 24 to 48 hours 

Static Charge 

Wiping electrodes 

Rinse electrode in 7.0 buffer and blot. Do no 
wipe electrode. 

Same readings in 
different buffers and 
samples 

Cracked or broken bulb 

Replace electrode. Use bulb guard. Avoid 
plunging electrode to bottom of container 
and spinning bars. Wetting cap will protect 
bulb between tests. 

Erratic LCD display 

Samples have low ionic strength 
(low sodium); e.g. distilled, de-
ionized, boiled, or lake water 
(high pressure) 

For each 50 ml of sample add 1 drop (50uL) of 
SAT. KCl (No alteration in pH will occur using 
inert KCl). 

Содержание PH220

Страница 1: ...USER MANUAL Waterproof Palm pH Meter Model PH220...

Страница 2: ...y does not include pH buffer temperature coefficient Note 2 Accuracy specified from 0 02pH to 13 98pH Temperature Compensation 32 to 194oF 0 to 90oC Automatic ATC Power 9 volt alkaline battery Display...

Страница 3: ...perature Note that there are other more specialized displays discussed later in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described in the Calibration s...

Страница 4: ...ibrated at least once each day pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automatical...

Страница 5: ...also display when the electrode is disconnected 8 When the test is complete rinse the electrode with distilled or deionized water Automatic Temperature Compensation ATC The millivolt output of a pH el...

Страница 6: ...the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and hold th...

Страница 7: ...the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagram 4 Repl...

Страница 8: ...60oC for 30 minutes Strong Alkaline measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1...

Страница 9: ...bration for most of our products Contact us for information on calibration and repair availability refer to the contact information below Annual calibrations should be performed to verify meter perfor...

Отзывы: