6

                                                   PH220-en-US_v3.5   3/21

 

 

Storing and Recalling Readings 

This instrument can store up to 25 measurement readings. Readings are saved even after the 
instrument has been turned off. 

STORE Readings 

1.

 

Take a measurement as described earlier in this manual. 

2.

 

Momentarily press the STORE/RECALL button to save the reading. The reading will be held 
on the display (the ‘HOLD’ and ‘STORE’ icons will appear). In addition, the storage location 
(1 to 25) will briefly appear on the lower LCD area. 

3.

 

Note: The temperature reading associated with the pH reading is also saved. 

4.

 

Press the STORE/RECALL button again to release the ‘hold’ feature. The instrument will 
then return to normal operation. 

5.

 

Note that when all 25 storage locations are filled the instrument begins writing over the 
previously stored data. 

RECALL Readings 

1.

 

Press and hold the STORE/RECALL button until a number appears on the main display. This 
number is the storage location for the most recent reading. 

2.

 

The number will switch off after 1 second and the stored reading will appear. The ‘RECALL’ 
icon will appear and remain on the entire time the recall mode is accessed. 

3.

 

While in the RECALL mode, use the STORE/RECALL button to scroll through the stored 
readings. The most recent readings will show first (LAST IN – FIRST OUT). 

4.

 

Press and hold the STORE/RECALL button until the ‘End’ display appears. Release the 
button and the instrument will return to the normal operation mode. 

5.

 

Note that if there are no readings in memory, the display will show ‘End’ when the user 
attempts to access the RECALL mode. 

ERASE (Clear) Readings 

1.

 

Press and hold STORE/RECALL and the CAL buttons simultaneously for 4 seconds. 

2.

 

The ‘CLr’ display will appear indicating that all readings have been erased. 

3.

 

Release the buttons to return to the normal operation mode. 

Temperature Offset Adjustment 

This instrument allows you to offset the temperature display by ±9

o

F (±5

o

C): 

1.

 

Press and hold the CALMODE, and STORE/RECALL buttons simultaneously for 6 seconds. 

2.

 

The main display will indicate the current offset. The lower display will show ‘t_O’. 

3.

 

Use the CAL button to decrease the offset value; use the MODE button to increase it. 

4.

 

Press and hold the STORE/RECALL button for 2 seconds to exit. The instrument will display 
‘SA’ (save) and ‘End’ before returning to the normal operation mode. 

5.

 

The temperature display will now indicate the actual temperature ± the offset value. 

Содержание PH220

Страница 1: ...USER MANUAL Waterproof Palm pH Meter Model PH220...

Страница 2: ...y does not include pH buffer temperature coefficient Note 2 Accuracy specified from 0 02pH to 13 98pH Temperature Compensation 32 to 194oF 0 to 90oC Automatic ATC Power 9 volt alkaline battery Display...

Страница 3: ...perature Note that there are other more specialized displays discussed later in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described in the Calibration s...

Страница 4: ...ibrated at least once each day pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automatical...

Страница 5: ...also display when the electrode is disconnected 8 When the test is complete rinse the electrode with distilled or deionized water Automatic Temperature Compensation ATC The millivolt output of a pH el...

Страница 6: ...the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and hold th...

Страница 7: ...the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagram 4 Repl...

Страница 8: ...60oC for 30 minutes Strong Alkaline measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1...

Страница 9: ...bration for most of our products Contact us for information on calibration and repair availability refer to the contact information below Annual calibrations should be performed to verify meter perfor...

Отзывы: